Changing built-in fuse
1.Ensure that the D1 unit is turned off and thathe power cable is not connected.
2.If the unit has been in use,wait five minutes to allow the unit to be fully discharged.
3.Pull out the Fuse Holder [14] from the unit and remove the old fuse.
4.Push the new fuse all the way into the Fuse Holder.Only use recommended fuse,see section Technical data.
5.Fit the Fuse Holder in place,by gently pushing until the holder snaps in place.