Maintaining the Equipment
3.Position the red
CAUTION: Use care with the IR filter. It is not secured in place. (For more information about the IR filter, see Replacing the (Square) IR Filter on page
4.Hold the red handle and pull the assembly straight out from the scanner.
5.Clean the filters with the
6.Replace the balance filters:
a.Remove the 3 screws.
b.Remove the filter retainer.
c.Remove the old filters and place the new filters in the filter holder.
The filters are coated on both sides. It is not important which side faces up or down.
d.Secure the filters with the retainer and the 3 screws.
7.Install the filter assembly in the scanner with the
8.Reinstall the lamp cover, tighten the thumbscrews, and select a filter position.
9.Calibrate the scanner (see Calibrating the Scanner on page