MegOhmMeter

Using the Meter

To measure insulation resistance, do the following:

1.Select the test voltage.

2. Connect the probes to the circuit to be measured.

The meter automatically detects if the circuit is energized, and displays the detectedWvoltage.

Warning

A repetitive beep and the flashing high voltage symbol (Y) warn the user if voltage more than 30 V ac or dc is present.

If the voltage is > 30 V ac or dc, remove the voltage from theacf05f.eps circuit under test before proceeding with the next step.

3.Push and hold the TEST button. The upper left display shows the test voltage applied to the circuit under test. The main display shows the resistance.

Note

The display shows - - - - until the TEST button is pushed.

13

Image 18
Contents 1520 Page Table of Contents Page List of Tables Fluke Unpacking the MeterFluke Safety Information and Symbols VDEFluke MegOhmMeter Key and Switch Descriptions Rotary switch Key functionsResistance Beeper function Low Resistance functionBacklight button Display Display DescriptionDisplay Using the Meter Connecting to the Circuit Under TestConnecting to the Circuit Under Test Auto-Shut-Off Measuring Insulation Resistance WDisplay shows - - until the Test button is pushed Fluke Using the Lock Function to Measure Insulation Resistance Measuring Low Resistance Using the Lock Function to Measure Resistance Measuring Resistance Measuring VoltageChecking the Battery Display ExampleCleaning Maintaining the MeterReplacing and Disposing of the BatteriesW AppearsFluke Replacing the Batteries Testing and Replacing the Fuse W Testing the FuseReplacing the Fuse Replacing the Fuse Optional Accessories Replacement Parts and Optional AccessoriesReplacement Part Part NumberService Centers Principle of Measurement for ResistanceSpecifications EnvironmentalMechanical Specifications Safety SpecificationsElectromagnetic Compatibility EMC Electrical Specifications Insulation ResistanceLo Ω Voltage ResistanceLimited Warranty & Limitation of Liability Netherlands Fluke