Fluke 1526 user manual Electrical Specifications, Insulation Resistance

Page 36

MegOhmMeter

Specifications

Electrical Specifications

Battery

C Size 1.5 V alkaline, ANSI/NEDA-14A, IEC- LR14 (4 pieces)

Fuse

 

6 mm x 32 mm (0.25 x 1.25 inch), 0.5 A 660 V,

 

 

Fast Acting, 50 A Minimum Interrupt rating

Insulation Resistance

Display Ranges

 

4.000

MΩ, 40.00 MΩ, 400.0 MΩ, 4000 MΩ

Measurement Range (per EN61557-2)

0.250

MΩ to 4000 MΩ

Resolution

0.001

MΩ to 10 MΩ

Accuracy

 

2 % + 2 counts

 

0.250

MΩ to 100.0 MΩ

 

 

10 % + 2 counts + 1 %/1000 MΩ

 

100.0

MΩ to 4000 MΩ

Analog Bar Graph

 

0 to

31

Image 36
Contents 1520 Page Table of Contents Page List of Tables Fluke Unpacking the MeterFluke Safety Information and Symbols VDEFluke MegOhmMeter Key and Switch Descriptions Rotary switch Key functionsResistance Beeper function Low Resistance functionBacklight button Display Display DescriptionDisplay Using the Meter Connecting to the Circuit Under TestConnecting to the Circuit Under Test Auto-Shut-Off Measuring Insulation Resistance WDisplay shows - - until the Test button is pushed Fluke Using the Lock Function to Measure Insulation Resistance Measuring Low Resistance Using the Lock Function to Measure Resistance Measuring Resistance Measuring VoltageChecking the Battery Display ExampleCleaning Maintaining the MeterReplacing and Disposing of the BatteriesW AppearsFluke Replacing the Batteries Testing and Replacing the Fuse W Testing the FuseReplacing the Fuse Replacing the Fuse Replacement Parts and Optional Accessories Replacement PartOptional Accessories Part NumberService Centers Principle of Measurement for ResistanceSpecifications EnvironmentalMechanical Specifications Safety SpecificationsElectromagnetic Compatibility EMC Electrical Specifications Insulation ResistanceLo Ω Voltage ResistanceLimited Warranty & Limitation of Liability Netherlands Fluke