Agilent Technologies 5964-8134 Front Panel Programming, Check the External Current Shunt Setting

Page 8

Check the External Current Shunt Setting

After you have completed your cable connections, you can turn on and check out the ac source as described in the Turn-On Checkout chapter in the ac source User’s Guide. One additional step in checking out a unit with Option 020/022 installed, is to make sure that the correct shunt value has been entered in the unit. To check the shunt value setting, proceed as follows:

Action

Display

1.Turn on the ac source and wait for the unit to finish its internal selftest.

2.

Press Shift Calibration and scroll to the SHUNT command.

SHUNT 0.01

3.Check to see if the resistance value on the display matches the value written on the outside of the external current shunt.

If it matches, press Meter to return to meter mode. You are done.

If it does not match, you must enter the correct value. Proceed as follows:

4.

Press Shift Calibration, scroll to CAL ON and press Enter. If necessary, enter

CAL ON

0.0

 

the calibration password from Entry keypad and press Enter. If the password is

 

 

 

correct the Cal annunciator will come on.

 

 

5.

Press Shift Calibration and scroll to the SHUNT command. Enter the shunt

SHUNT

0.01

 

value written on the outside of the external current shunt exactly as it is written on

 

 

 

the label. Then press Enter.

 

 

6.

Press Shift Calibration, scroll to CAL SAVE, and press Enter.

CAL:SAVE

7.

Press Shift Calibration, select CAL OFF, and press Enter to exit Calibration.

CAL OFF

NOTE: The shunt value that you enter applies only to the Power Analyzer input and has no effect on ac source output measurements. Therefore, ignore the φ1 and φ2 annunciators while entering the shunt value.

Front Panel Programming

Setting the Output Frequency Modulation

To set the output frequency modulation:

 

Action

Display

 

1.

On the Function keypad, press Frequency. Then scroll to the FM:FREQ command

FM:FREQ

1

 

to enter a modulating frequency. Use the numeric keypad to enter a modulating value

 

 

 

and press Enter.

 

 

2.

To set the peak frequency deviation of the output frequency modulation, press

FM:DEV

10

 

Frequency. Then scroll to the FM:DEV command to enter a frequency deviation.

 

 

 

Use the numeric keypad to enter a value and press Enter.

 

 

 

The deviation frequency must be less than the programmed output frequency by at

 

 

 

least 0.001 Hz. Otherwise, it will result in Error 613.

 

 

3.

To enable frequency modulation, press Frequency, scroll to FM OFF, select the

FM ON

 

 

ON parameter and press Enter.

 

 

8

Image 8
Contents Addendum Contents Agilent Dual Power Analyzer Description Agilent Option 022 DescriptionScpi Programming Commands Additions Front Panel Menus AdditionsInstallation Power Analyzer input. Connector plug is removableTurn the unit off before connecting any wires Rear Panel ConnectionsMeasurement Cable Connections External Shunt Line and Load Connections Check the External Current Shunt Setting Setting the Output Frequency ModulationFront Panel Programming To set the output frequency modulationUsing the Front Panel Display with the Power Analyzer Input Making Power Analyzer Input MeasurementsTo select the measurement source Use the Input menu for selecting the frequency sourceAdditional Measurement Commands MEASureTVOLt? FETChTVOLt?MEASureTVOLtABSolute? FETChTVOLtABSolute? MEASureVOLTageRANKed? FETChVOLTageRANKed?Query Syntax Parameters Examples Parameters ExamplesSENSeFREQuencySOURce Additional Sense CommandsSENSeNSELect Additional Frequency Modulation Commands FMDEViationFMFREQuency RST Value UnitAdditional Calibration Commands Command Syntax Parameters ExamplesCALibrateVOLTageEXTernal CALibrateCURRentEXTernalBehavior Differences of Existing Scpi Commands Programming Example UPS Transfer Time MeasurementPrograms Agilent AC Source AC Source GUI Using the Power Analyzer Input with the GUIMaking a Measurement Making a Voltage Phase Difference Measurement Performing a Transfer Time Test Modulating the Output FrequencyPower Analyzer Input Specifications Table A-1. Performance SpecificationsTable A-2. Supplemental Characteristics Table B-1. Equipment Required Verification and Calibration for Option 020/022Equipment Required Verification Procedure Test Set upPower Analyzer Voltage Measurement Accuracy Power Analyzer Current Measurement AccuracyCalibration Procedure Enable Calibration ModeDisplay Command Function Summary of Front Panel Calibration MenuCAL Save

5964-8134 specifications

Agilent Technologies 5964-8134 is an advanced analytical instrument designed for versatile applications in laboratory settings. It serves as a crucial component for professionals working in fields such as forensic science, environmental testing, and pharmaceutical analysis.

One of the primary features of the 5964-8134 is its exceptional chromatographic performance. The device employs sophisticated separation techniques, utilizing a high-efficiency column that enhances the resolution of complex mixtures. This capability allows scientists to identify and quantify trace levels of compounds in samples with high precision.

Equipped with a state-of-the-art detector, the 5964-8134 offers a broad dynamic range to accommodate various analyte concentrations. This feature ensures that both minor and major components can be accurately measured, making it ideal for applications involving diverse sample matrices. The integrated software provides users with easy access to data analysis tools, enabling seamless interpretation of results.

The 5964-8134 is built on cutting-edge technology, including advanced mass spectrometry which enhances its sensitivity and specificity. This technology enables the detection of low-abundance compounds in complex samples, providing insights that would be difficult to obtain using traditional methods. Additionally, the instrument supports multiple ionization techniques, allowing users to tailor the analysis to specific requirements of their project.

Another characteristic of the Agilent 5964-8134 is its user-friendly interface. The equipment is designed with the operator in mind, featuring intuitive controls and a touchscreen display that promotes ease of use. This design minimizes the learning curve for new users, enabling efficient operation and streamlined workflow in busy laboratory environments.

Moreover, the 5964-8134 ensures robust performance with its advanced temperature control and reliable fluidic systems. These features contribute to improved reproducibility and stability during analyses, which are critical in achieving consistent results across experiments.

With its combination of high performance, cutting-edge technologies, and user-centric design, Agilent Technologies 5964-8134 stands out as a leading choice for professionals seeking reliable and accurate analytical solutions. As laboratories continue to demand innovative instruments that push the boundaries of research capabilities, the 5964-8134 proves to be an essential tool in scientific discovery and analysis.