Calibrating and Entering Power Analyzer Voltage Calibration Values

Action

Display

4.Connect an Agilent 3458A DVM (in synchronous ac volts mode) to the ac source output terminals. Short the Power Analyzer voltage terminals (V SENSE) to ensure no dc offset error.

5.

Press Shift Calibration, scroll to CAL VOLT EXT, and press Enter.

CAL:VOLT:EXT

6.Leave the ac source on and remove the short from the Power Analyzer voltage terminals.

7.Connect the Power Analyzer voltage sense terminals to the Main Output terminals of the ac source (see figure 3-1A).

8.

Press Shift Calibration, scroll to CAL LEV P1 and press Enter.

CAL:LEV P1

 

The ac source will output 300 Vac 60 Hz sine wave.

 

9.

Press Shift Calibration, scroll to CAL DATA, and use the Entry keypad to

CAL:DATA 0.00

 

enter the voltage value displayed on the Agilent 3458A DVM.

 

10.

Press Shift Calibration, scroll to CAL SAVE and press Enter.

CAL SAVE

Calibrating and Entering Power Analyzer Current Calibration Values

 

Action

Display

11.Short the Power Analyzer current sense terminals (I SENSE) to ensure no dc offset error.

12. Press Shift Calibration, scroll to CAL CURR EXT and press Enter.

CAL:CURR:EXT

13.Leave the ac source on and remove the short from the Power Analyzer current terminals.

14Connect the Power Analyzer current terminals to the output of the function generator (see figure 3-1B). Set the function generator output to 50 mV, 60 Hz.

15.Connect the Power Analyzer current terminals to the Agilent 3458A input terminals. Note that the Agilent 3458A must be in synchronous ac volts mode and a 1.0 uF capacitor must be installed across the Agilent 3458A input terminals.

16.

Press Shift Calibration, scroll to CAL LEV P1, and press Enter.

CAL LEV P1

17.

Press Shift Calibration and scroll to CAL DATA. Use the Entry keypad to

CAL:DATA 0.00

 

enter the voltage value displayed on the Agilent 3458A DVM.

 

18

Change the function generator output setting to 500 mV.

 

19.

Press Shift Calibration, scroll to CAL LEV P1 command, use ° to scroll to

CAL:LEV P2

 

the P2 parameter, and press Enter.

 

20.

Press Shift Calibration and scroll to the CAL DATA. Use the Entry keypad to

CAL:DATA 0.00

 

enter the voltage value displayed on the Agilent 3458A DVM.

 

21

Press Shift Calibration, scroll to CAL SAVE and press Enter

CAL SAVE

26

Image 26
Contents Addendum Contents Agilent Dual Power Analyzer Description Agilent Option 022 DescriptionScpi Programming Commands Additions Front Panel Menus AdditionsTurn the unit off before connecting any wires InstallationPower Analyzer input. Connector plug is removable Rear Panel ConnectionsMeasurement Cable Connections External Shunt Line and Load Connections Front Panel Programming Check the External Current Shunt SettingSetting the Output Frequency Modulation To set the output frequency modulationTo select the measurement source Using the Front Panel Display with the Power Analyzer InputMaking Power Analyzer Input Measurements Use the Input menu for selecting the frequency sourceAdditional Measurement Commands MEASureTVOLt? FETChTVOLt?Query Syntax Parameters Examples MEASureTVOLtABSolute? FETChTVOLtABSolute?MEASureVOLTageRANKed? FETChVOLTageRANKed? Parameters ExamplesSENSeFREQuencySOURce Additional Sense CommandsSENSeNSELect FMFREQuency Additional Frequency Modulation CommandsFMDEViation RST Value UnitCALibrateVOLTageEXTernal Additional Calibration CommandsCommand Syntax Parameters Examples CALibrateCURRentEXTernalBehavior Differences of Existing Scpi Commands Programming Example UPS Transfer Time MeasurementPrograms Agilent AC Source AC Source GUI Using the Power Analyzer Input with the GUIMaking a Measurement Making a Voltage Phase Difference Measurement Performing a Transfer Time Test Modulating the Output FrequencyPower Analyzer Input Specifications Table A-1. Performance SpecificationsTable A-2. Supplemental Characteristics Table B-1. Equipment Required Verification and Calibration for Option 020/022Equipment Required Verification Procedure Test Set upPower Analyzer Voltage Measurement Accuracy Power Analyzer Current Measurement AccuracyDisplay Command Function Calibration ProcedureEnable Calibration Mode Summary of Front Panel Calibration MenuCAL Save

5964-8134 specifications

Agilent Technologies 5964-8134 is an advanced analytical instrument designed for versatile applications in laboratory settings. It serves as a crucial component for professionals working in fields such as forensic science, environmental testing, and pharmaceutical analysis.

One of the primary features of the 5964-8134 is its exceptional chromatographic performance. The device employs sophisticated separation techniques, utilizing a high-efficiency column that enhances the resolution of complex mixtures. This capability allows scientists to identify and quantify trace levels of compounds in samples with high precision.

Equipped with a state-of-the-art detector, the 5964-8134 offers a broad dynamic range to accommodate various analyte concentrations. This feature ensures that both minor and major components can be accurately measured, making it ideal for applications involving diverse sample matrices. The integrated software provides users with easy access to data analysis tools, enabling seamless interpretation of results.

The 5964-8134 is built on cutting-edge technology, including advanced mass spectrometry which enhances its sensitivity and specificity. This technology enables the detection of low-abundance compounds in complex samples, providing insights that would be difficult to obtain using traditional methods. Additionally, the instrument supports multiple ionization techniques, allowing users to tailor the analysis to specific requirements of their project.

Another characteristic of the Agilent 5964-8134 is its user-friendly interface. The equipment is designed with the operator in mind, featuring intuitive controls and a touchscreen display that promotes ease of use. This design minimizes the learning curve for new users, enabling efficient operation and streamlined workflow in busy laboratory environments.

Moreover, the 5964-8134 ensures robust performance with its advanced temperature control and reliable fluidic systems. These features contribute to improved reproducibility and stability during analyses, which are critical in achieving consistent results across experiments.

With its combination of high performance, cutting-edge technologies, and user-centric design, Agilent Technologies 5964-8134 stands out as a leading choice for professionals seeking reliable and accurate analytical solutions. As laboratories continue to demand innovative instruments that push the boundaries of research capabilities, the 5964-8134 proves to be an essential tool in scientific discovery and analysis.