Agilent Technologies 5964-8134 manual Verification and Calibration for Option 020/022

Page 22

Verification and Calibration for Option 020/022

This section includes verification and calibration procedures for the Agilent Dual Power Analyzer Option 020/022. Add these procedures to those described in Appendix B of the AC Source User’s Guide.

The verification procedures do not check all the operating parameters, but verify that the ac source is performing properly. Performance Tests, which check all the specifications of the ac source, are given in the applicable ac source Service Manual.

Because the output of the ac source must be enabled during verification or calibration, proceed with caution, since voltages and currents will be active at the output terminals.

Important Perform the verification procedures before calibrating your ac source. If the ac source passes the verification procedures, the unit is operating within its calibration limits and does not need to be re-calibrated.

WARNING LETHAL VOLTAGES. Ac sources can supply 424 V peak at their output. DEATH on contact may result if the output terminals or circuits connected to the output are touched when power is applied. These procedures must be performed by a qualified electronics technician or engineer trained on this equipment.

Equipment Required

The equipment listed in the following table, or the equivalent to this equipment, is required for verification and calibration.

Table B-1. Equipment Required

Equipment

5-pin connector plug

Digital Voltmeter

Current Monitor1

Load Resistor

Capacitor (calibration only)

Function Generator

(calibration only)

GPIB Controller

Characteristics

This connector is supplied with the ac source when ordered with Option 020/022.

Resolution: 10 nV @ 1 V

Readout: 8.5 digits

Accuracy: >20 ppm

0.01 Ω, ±200 ppm,

30 ohm, 100 Watts minimum

1.0 μF, 50 V

50 mV, 500 mV, 60 Hz sinewave capability

Full GPIB capabilities

Recommended Model

0360-2681

Agilent 3458A

Guildline 7320/0.01

0160-3490

Agilent 33120A

HP Series 200/300 or equivalent

1The 4- terminal current shunt is used to eliminate output current measurement error caused by voltage drops in the load leads and connections. Connect the voltmeter directly to these current-monitoring terminals.

22

Image 22
Contents Addendum Contents Agilent Dual Power Analyzer Description Agilent Option 022 DescriptionScpi Programming Commands Additions Front Panel Menus AdditionsTurn the unit off before connecting any wires InstallationPower Analyzer input. Connector plug is removable Rear Panel ConnectionsMeasurement Cable Connections External Shunt Line and Load Connections Front Panel Programming Check the External Current Shunt SettingSetting the Output Frequency Modulation To set the output frequency modulationTo select the measurement source Using the Front Panel Display with the Power Analyzer InputMaking Power Analyzer Input Measurements Use the Input menu for selecting the frequency sourceAdditional Measurement Commands MEASureTVOLt? FETChTVOLt?Query Syntax Parameters Examples MEASureTVOLtABSolute? FETChTVOLtABSolute?MEASureVOLTageRANKed? FETChVOLTageRANKed? Parameters ExamplesSENSeNSELect Additional Sense CommandsSENSeFREQuencySOURce FMFREQuency Additional Frequency Modulation CommandsFMDEViation RST Value UnitCALibrateVOLTageEXTernal Additional Calibration CommandsCommand Syntax Parameters Examples CALibrateCURRentEXTernalBehavior Differences of Existing Scpi Commands Programming Example UPS Transfer Time MeasurementMaking a Measurement Using the Power Analyzer Input with the GUIPrograms Agilent AC Source AC Source GUI Making a Voltage Phase Difference Measurement Performing a Transfer Time Test Modulating the Output FrequencyPower Analyzer Input Specifications Table A-1. Performance SpecificationsTable A-2. Supplemental Characteristics Equipment Required Verification and Calibration for Option 020/022Table B-1. Equipment Required Verification Procedure Test Set upPower Analyzer Voltage Measurement Accuracy Power Analyzer Current Measurement AccuracyDisplay Command Function Calibration ProcedureEnable Calibration Mode Summary of Front Panel Calibration MenuCAL Save

5964-8134 specifications

Agilent Technologies 5964-8134 is an advanced analytical instrument designed for versatile applications in laboratory settings. It serves as a crucial component for professionals working in fields such as forensic science, environmental testing, and pharmaceutical analysis.

One of the primary features of the 5964-8134 is its exceptional chromatographic performance. The device employs sophisticated separation techniques, utilizing a high-efficiency column that enhances the resolution of complex mixtures. This capability allows scientists to identify and quantify trace levels of compounds in samples with high precision.

Equipped with a state-of-the-art detector, the 5964-8134 offers a broad dynamic range to accommodate various analyte concentrations. This feature ensures that both minor and major components can be accurately measured, making it ideal for applications involving diverse sample matrices. The integrated software provides users with easy access to data analysis tools, enabling seamless interpretation of results.

The 5964-8134 is built on cutting-edge technology, including advanced mass spectrometry which enhances its sensitivity and specificity. This technology enables the detection of low-abundance compounds in complex samples, providing insights that would be difficult to obtain using traditional methods. Additionally, the instrument supports multiple ionization techniques, allowing users to tailor the analysis to specific requirements of their project.

Another characteristic of the Agilent 5964-8134 is its user-friendly interface. The equipment is designed with the operator in mind, featuring intuitive controls and a touchscreen display that promotes ease of use. This design minimizes the learning curve for new users, enabling efficient operation and streamlined workflow in busy laboratory environments.

Moreover, the 5964-8134 ensures robust performance with its advanced temperature control and reliable fluidic systems. These features contribute to improved reproducibility and stability during analyses, which are critical in achieving consistent results across experiments.

With its combination of high performance, cutting-edge technologies, and user-centric design, Agilent Technologies 5964-8134 stands out as a leading choice for professionals seeking reliable and accurate analytical solutions. As laboratories continue to demand innovative instruments that push the boundaries of research capabilities, the 5964-8134 proves to be an essential tool in scientific discovery and analysis.