Teledyne HFM-E-200 instruction manual 0014 Q, HFC-E-202 Orifice

Page 18

Where:

 

 

 

 

 

Formula 1:

Formula 2:

D =

0.0028σQ

 

D =

0.0014Q

Pu

ΔΡ • Ρd

 

 

 

 

 

 

σ

D = Diameter in inches

Q = Flow rate in standard liters per minute

P = Pu - Pd in PSI

Pu = Upstream pressure in PSIA

Pd = Downstream pressure in PSIA

σ= Specific gravity of gas

Choose the orifice form Section 5.0 that has the closes larger diameter to the calculated diameter.

4.7.1. HFC-E-202 Orifice

To change the orifice in the HFC-E-202 series, turn the instrument upside-down and turn the orifice counterclockwise with a 9/64” Allen wrench until it stops coming out. Then grasp the exposed orifice end and pull it straight out. See Figure 4.2.

Prior to reinstallation of the orifice, inspect the two O-rings mounted on it for damage. Replace if cut or gouged.

Lubricate the O-rings slightly with a silicone based grease, and the threads with anti-galling compound. Push the orifice into its hole and screw it in until it is flush with the instrument base. Apply pressure to the inlet side of the instrument.

Turn the Valve Override switch to CLOSE or unplug the instrument. Screw the orifice in a few more turns until the flow through the instrument stops, then turn it an additional 1/8 turn clockwise.

PARKER O-RING

#2-008 VITON

ORIFICE

(see P/L per P/#

Fig 4.2

HFM-E-200/HFC-E-202

Page 18 of 30

Image 18
Contents Teledyne Hastings Manual Print History Table of Contents Installation and Operation FeaturesSpecifications Swagelok, 1/8 Swagelok, VCR, VCOInterconnecting Cables Other AccessoriesHastings Power supplies Optional 4-20 mA Current OutputReceiving Inspection Power RequirementsOutput Signal Mechanical ConnectionsElectrical Connections Operation Operating ConditionsZero Check Operation with External Devices Operation with a Hastings power supplyOperation with a power supply other than a Hastings High Pressure OperationRange Changing Operation with an external sensor. FigSoft Start Sensor Theory of OperationOverall Functional Description ElectronicsShunt Valve Maintenance Authorized MaintenanceTroubleshooting Adjustments Calibration Procedure FigureMiscellaneous adjustments Printed Circuit Board ReplacementSensor Replacement Inlet Removal0014 Q HFC-E-202 OrificeReplacement Parts Conversion Factor Table BF3C2H7N C3H8O SO2 Warranty Repair Policy Non-Warranty Repair PolicyDrawings HFM-E-200/HFC-E-202 HFM-E-200/HFC-E-202 HFM-E-200/HFC-E-202 HFM-E-200/HFC-E-202 HFM-E-200/HFC-E-202

HFC-E-202, HFM-E-200 specifications

Teledyne HFC-E-202 and HFM-E-200 are advanced flow meters designed to provide precise measurement and monitoring of gas and liquid flows in various industrial applications. Both devices are engineered to meet the demands of sectors such as petrochemicals, pharmaceuticals, and food and beverage, where accuracy and reliability are paramount.

The Teledyne HFC-E-202 is a highly sensitive flow meter that utilizes a thermal mass flow measurement technology. This technology enables the HFC-E-202 to accurately measure the mass flow rate of gases regardless of temperature and pressure conditions. One of its key features is its ability to provide high precision and fast response times, making it suitable for applications that require immediate data feedback. The device is also equipped with advanced diagnostics that allow users to monitor the health of the system and identify potential issues before they escalate.

On the other hand, the HFM-E-200 is particularly renowned for its versatility in measuring both gas and liquid flows. It employs differential pressure technology, which is highly effective in a wide range of operating conditions. The HFM-E-200 incorporates a robust design that can withstand harsh environments, ensuring longevity and reliability. Its user-friendly interface allows for easy configuration and monitoring, catering to both novice and experienced technicians.

Both flow meters feature digital output options, facilitating seamless integration with existing control systems. This enables real-time data acquisition and monitoring, enhancing process control and efficiency. Furthermore, their compact design minimizes installation space, allowing for easy placement in tight operational areas.

In terms of safety and compliance, both the HFC-E-202 and HFM-E-200 are designed to meet strict industry standards, ensuring they can be deployed in critical applications without compromising safety.

In conclusion, Teledyne HFC-E-202 and HFM-E-200 flow meters stand out due to their innovative technologies, accuracy, and durability. Their unique features position them as essential tools for a variety of industrial sectors, enhancing operational efficiency and ensuring reliable fluid flow measurement. With Teledyne’s commitment to quality and innovation, users can trust that these devices will meet their demands both now and in the future.