Omega Engineering FLR6000 series, FLR7000 series manual Correction Factors, Fluid Selection

Page 18

FLR Flow Transmitter Installation & Programming Instructions

Application Information – Pneumatic

NOTE: Pressure and temperature readings must be taken at the flow transmitter inlet to ensure accurate correction factors.

The pneumatic flow transmitter is calibrated for air in standard cubic feet per minute (scfm) at 1.0 s.g. (70°F

@100 psi), and liters per second (lps) at 1.0 s.g.

(21°C @ 6.9 bar).

Figure 10. System Schematic

Table 4 - Correction Factors

FLOW RATE CORRECTION FACTORS FOR GAS

SCFM (Actual) =

SCFM (Indicated) f1

= conversion

factor

for inlet pressure

 

f1 x f2 x f3

 

f2 = conversion factor for temperature

 

 

 

 

 

f3 = conversion factor for specific gravity

 

 

 

 

 

 

 

 

 

 

PRESSURE CORRECTION FACTOR (f1)

 

 

 

 

 

 

 

 

Operating Pressure (psig)

 

 

 

psig

25

50

75

100

125

150

175

200

225

250

f1

1.700

1.331

1.131

1.00

.902

.835

.778

.731

.692

.658

 

 

 

 

f1 =

 

114.7

 

 

 

 

 

 

 

 

14.7+psig

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TEMPERATURE CORRECTION FACTOR (f2)

 

°F

10

30

50

70

90

110

 

130

150

170 190

f2

.942

.962

.981

1.00

1.018

1.037

1.055

1.072

1.090 1.107

Table 5 - Fluid Selection

 

 

Specific Gravity

Correction Factor Std. Scale

Aluminum

Brass

T316 SSt

T303 SSt

Viton

EPR

Polycarbonate

Nylon

Pyrex™

 

Air / Gas

 

Air

1.0

1.000

R

R

R

R

R

R

R

R

R

 

Argon (A)

1.38

1.175

R

R

R

R

R

R

R

R

R

 

CarbonDioxide(CO2)

1.53

1.237

R

R

R

R

R

R

R

R

R

 

Freon 11 (CCI3F)

4.92

2.218

R

R

R

R

R

R

R

R

R

 

Freon 12 (CCI2F)

4.26

2.060

R

R

R

R

R

R

R

R

R

 

Helium (HE)

0.14

0.374

R

R

R

R

R

R

R

R

R

 

Hydrogen (H2)

0.07

0.265

R

R

R

R

R

R

R

R

R

 

Natural Gas

0.60

0.775

C

C

R

C

R

N

C

R

R

 

Nitrogen (N2)

0.97

0.985

C

C

R

R

R

R

C

R

R

 

Oxygen (O2)

1.10

1.049

R

R

R

R

R

R

R

R

R

 

Propane (C3H8)

1.57

1.253

R

R

R

R

R

N

N

R

R

f 2 =

460 + °F

530

R = Recommended

N = Not recommended C = Consult Factory

SPECIFIC GRAVITY CORRECTION FACTOR (f3)

f3 = Specific Gravity

Note: Table 4 is included to show the correction algorithms included in the program to perform pressure, temperature, and specific gravity corrections. When configuring the FLR Flow Transmitter, enter the actual operating pressure, temperature, and specific gravity values, not the correction factors.

Page 16

Image 18
Contents User’s Guide Servicing North America Introduction FLR Flow TransmitterII. Specifications III. Installation Electrical Connections Terminology IV. Operation Installing the TransmitterNormal Operation RUN Mode Operating the MeterProgramming Operation Program Mode Cover Removal/ReinstallationList Item Selection Procedure Programming Descriptions Total Flow UnitsFLR Flow Transmitter Installation & Programming Instructions FLR Flow Transmitter Installation & Programming Instructions Meter Size Types Zero CaptureOutput Mode Maintenance Cartridge Cleaning on page 5 and onVI. Troubleshooting VII. AppendixCorrection Factors Fluid SelectionFlow vs. Pressure Drop Water Petroleum Fluids Phosphate Ester I. Oil Air / Compressed Gases FLR Flow Transmitter Installation & Programming Instructions FLR Flow Transmitter Installation & Programming Instructions WARRANTY/ Disclaimer Shop online at omega.com