Omega Engineering FLR9000 series, FLR7000 series manual Programming Descriptions, Total Flow Units

Page 11

FLR Flow Transmitter Installation & Programming Instructions

Programming Flow Chart

The programming flow chart on pages 10 and 11 will aid understanding of the menu structure of the FLR Flow Transmitter. It will also help with understanding the available configuration selections.

Programming Descriptions

Display Mode

The meter can display RATE (flow rate) or TOTAL (total accumulated flow) or alternate between BOTH rate and total. Its displayed name is DISPLAY and is viewed or changed using the List Item Selection Procedure found on page 8.

Rate Units of Measure

The meter allows the selection of many common rate units. Its displayed name is RATE UNT and is viewed or changed using the List Item Selection Procedure found on page 8.

Rate (Time) Interval

The meter allows selection of several intervals based on time. Its displayed name is RATE INT and is viewed or changed using the List Item Selection Procedure found on page 8.

Total Units of Measure

If the total flow is being displayed, the units for the total must first be chosen. The monitor allows the choice of many common totalization units. Its displayed name is TOTL UNT and is viewed or changed using the List Item Selection Procedure found on page 8.

Total Display Multiplier

The meter has the ability to accumulate the flow total in multiples of ten. For example, if the most desirable totalization unit is 1,000 gallons, the monitor can easily be set up for this requirement. Once back in RUN mode every time the total display increments by one digit the actual total would be an additional 1,000 gal- lons. At 1,000 total gallons the total display would read 1, at 3,000 gallons the total display would read 3, etc. This feature allows the unit to accumulate totals that would exceed the 8 digit display capacity. Table 2 lists the available selection choices. Its displayed name is TOTL EXP and is viewed or changed using the List Item Selection Procedure found on page 8.

Table 2 - Total Flow Units

E-2

Total display number indicates

increments of 0.01 unit

 

 

 

E-1

Total display number indicates

increments of 0.1 unit

 

 

 

E0

Total display number indicates

increments of 1 unit

 

 

 

E1

Total display number indicates

increments of 10 units

 

 

 

E2

Total display number indicates

increments of 100 units

 

 

 

E3

Total display number indicates

increments of 1,000 units

 

 

 

E4

Total display number indicates

increments of 10,000 units

 

 

 

E5

Total display number indicates

increments of 100,000 units

 

 

 

E6

Total display number indicates

increments of 1,000,000 units

 

 

 

Scale Factor

The scale factor is used to span the meter. Its dis- played name is SCALE F and is viewed or changed using the Numeric Value Entry Procedure found on page 8.

Page 9

Image 11
Contents User’s Guide Servicing North America FLR Flow Transmitter IntroductionII. Specifications III. Installation Electrical Connections Terminology Operating the Meter Installing the TransmitterNormal Operation RUN Mode IV. OperationCover Removal/Reinstallation Programming Operation Program ModeList Item Selection Procedure Total Flow Units Programming DescriptionsFLR Flow Transmitter Installation & Programming Instructions FLR Flow Transmitter Installation & Programming Instructions Zero Capture Meter Size TypesOutput Mode Cartridge Cleaning on page 5 and on MaintenanceVII. Appendix VI. TroubleshootingFluid Selection Correction FactorsFlow vs. Pressure Drop Water Petroleum Fluids Phosphate Ester I. Oil Air / Compressed Gases FLR Flow Transmitter Installation & Programming Instructions FLR Flow Transmitter Installation & Programming Instructions WARRANTY/ Disclaimer Shop online at omega.com