Inflating the cuff: Close the bulb’s air valve by turning the air release valve clockwise. Squeeze the inflation bulb at a steady rate until the gauge’s needle points at approximately 30mmHg above the individual’s normal systolic pressure value. If the individual’s normal blood pressure is not known, it is recommended to inflate to 200mmHg.
Systolic blood pressure reading: Open the air release valve slowly by turning it
Diastolic blood pressure reading: Allow the pressure to continue dropping at the same deflation rate. Note the reading on the gauge when the last audible thumping, swishing, or blowing sound is heard; this is the diastolic blood pressure reading. After a few seconds have passed and no audible thumping, swishing, or blowing sounds are heard, deflate the cuff using the air release valve completely. Remove the cuff and stethoscope from the arm.
Record the systolic and diastolic readings. Repeat the measurement two or more times to ensure accuracy. Only a patient’s physician is qualified to analyze blood pressure.
MAINTENANCE
Recommended care and maintenance:
Do not drop or pull excessively on the blood pressure unit or stethoscope components.
Never inflate beyond 300mmHg.
Do not expose this product to direct sunlight.
Do not place this product in contact with sharp objects which could pierce the material and cause damage.
Do not dismantle or disassemble this product.
Cleaning | Wipe off the blood pressure unit, stethoscope and bulb with a clean, |
| damp cloth. |
|
|
| Wash the cuff with soap and cold water, rinse and air dry. |
|
|
Storage | Always deflate the cuff completely before storage. |
|
|
| Store the complete instrument in the provided storage case. |
|
|
Lumiscope Designer Professional Combo Kit with | 3 |