Technical specifications
curverebreathing. As a result of manufacturing variations, the vent flow rate may vary.
| 70 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| Pressure | Flow |
| 60 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
(L/min) |
|
|
|
|
|
|
|
|
|
|
|
|
|
| (cm H2O) | (L/min) | |
50 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
40 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 2 | 14 | |
rate |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
30 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 9 | 31 | |
flow |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
20 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 16 | 43 | |
Vent | 10 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 23 | 53 | |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
| 2 | 4 | 6 | 8 | 10 | 12 | 14 | 16 | 18 | 20 | 22 | 24 | 26 | 28 | 30 | ||
|
|
|
|
|
| Mask pressure (cm H2O) |
|
|
|
|
| 30 | 61 | ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
Dead space | Physical dead space is the empty volume of the mask |
information | to the end of the swivel. |
| It is 101.7 mL. |
|
|
Therapy pressure | 2 to 30 cm H2O |
Compliance Mask compliance between 5 cm H2O and 15 cm H2O is less than 1.57 mL/cm H2O
Resistance | Drop in pressure measured (nominal) |
| at 50 L/min: 0.8 cm H2O |
| at 100 L/min: 3.1 cm H2O |
Sound | DECLARED |
| VALUES in accordance with ISO 4871. The |
| sound power level of the mask is 40.6 dBA, with |
| uncertainty 4 dBA. The |
| level of the mask at a distance of 1 m is 32.6 dBA, with |
| uncertainty 4 dBA. |
Environmental Operating temperature: +5°C to +40°C
conditions Operating humidity: 15% to 95%
Storage and transport humidity: up to 95% non- condensing.
Gross | Mask fully assembled with elbow assembly – no |
dimensions | headgear. |
| 35 mm (H) x 430 mm (W) x 40 mm (D) |
|
|
18