Omega Engineering OS533E manual Adjusting Emissivity, Using the Lock Function

Page 26
2-11

Using the Handheld Infrared Thermometer 2

2.3.1 Adjusting Emissivity

°F

Refer to Appendices B and C for information on emissivity.

1.Determine the emissivity of the target.

2.Aim at the target and pull the trigger.

3.If necessary, press the key to increment the target emissivity or press the C F key to decrement the target emissivity.

NOTE

The Emissivity Display Mode (E) appears every time the trigger is pulled regardless of how the Display Mode was previously set.

NOTE

The emissivity setting does not change when the thermometer is turned off. However, when the batteries are replaced, the emissivity is reset to 0.95, the default value.

2.3.2 Using the LOCK Function

This function electronically locks the trigger mechanism:

1.Pull the trigger and press the LOCK key to lock the trigger in the Emissivity

and Distance Display Mode. The icon will appear on the display.

2.Release the trigger. This allows the thermometer to operate continuously whether or not the trigger is pulled.

NOTE

To unlock the trigger function, while in Emissivity and

Distance Display Modes press the LOCK key again, and

the icon is no longer displayed.

The LOCK key also enables/disables alarm functions, and resets calculated temperature values (MAX, MIN, dIF, AVG).

2.3.3 Using the Trigger Function

Besides turning on the thermometer by pulling the trigger, you can lock the trigger electronically by pressing the trigger button twice.

The icon is displayed. You can unlock the trigger by pressing the trigger button twice again.

Image 26
Contents User’sGuide Servicing North America Unpacking Instructions Accessories Page Unpacking Instructions Chapter General Description Chapter Maintenance General Description Introduction OS530LE OS530HRE OS530LE-CF OS533E-CF OS534E-CF Features OS523E OS524E Parts of the Thermometer Front of the Thermometer LCDKey General Description General Description Ac Power Operation Installing the BatteriesOperating the Thermometer Field of View PositionsField of View OS534E, OS523E-1 Field of View OS53xE-CF Field of View OS523E-3 Measurement Techniques 11 Recorder Hookup Real Time Mode Active Operation 12. General Operational Block DiagramReal Time Modes 13. Visual Function Flow Chart Using the Trigger Function Adjusting EmissivityUsing the Lock Function Using the Distance Function Must be hard, flat, and reflective to ultrasonic pulseDS = 17. Field of View of Distance Meter HH-DM Calculating Temperature Values Laser Sighting StatusThermocouple Input OS532E, OS533E, OS534E Changing the Temperature from F to C or vice VersaTurning on the Display Backlighting Using the Alarm Functions To set the low alarm value OS533E, OS534E, OS523E, OS524E Use the Ambient Target Temperature Compensation Atcf PC User Application, OS530 Series Operation19. Main Menu Settings Menus 10. Settings Menu21. Typical Temperature Data File PC Interface Commands AsciiString Description Storing Temperature Data on Command OS534E, OS523E, OS524E Logging Temperature Data in Real Time OS523E, OS524E,OS534E Erasing the Temperature Data from Memory Recall Mode Passive Operation ModeRecall Modes Reviewing the Last Parameters Laser Sighting Description Right Side ViewTwo Laser Configurations Operating the Laser SightingLines of Sight of the Laser Sighting and Thermometer Installing and Operating the Sighting Scope Sighting ScopeInstalling the Sighting Scope Safety Warning Maintenance Replacing the BatteriesServicing the Laser Sighting Cleaning the LensCalibrating the Thermometer Troubleshooting Guide Problem SolutionTroubleshooting Guide Problem Troubleshooting Guide Specifications are for all models except where noted SpecificationsGND With Built-in Distance Module Laser Sighting Distance Measuring Built-in-DM Specifications Glossary of Key Strokes Glossary of Key Strokes Thermal Radiation Appendix How Infrared Thermometry Works aBlackbody Calculating Temperature Optics Field of View Metals Appendix Emissivity ValuesNonmetals Appendix Determining an Unknown Emissivity Figure C-1. Determining Emissivity Figure C-2. Determining Emissivity with a Drilled Hole Appendix Determining an Unknown Emissivity Index Field of View Diagrams Recall WARRANTY/DISCLAIMER Temperature PH/CONDUCTIVITY