Omega Engineering OS533E manual Specifications are for all models except where noted

Page 60

Specifications 7

(Specifications are for all models except where noted)

THERMOMETER

Measuring

OS530HRE,

-30°C to 121°C (-22°F to 250°F)

Temperature

OS530LE,

 

Range:

OS533E, OS532E:

-23°C to 538°C (-10°F to 1000°F)

 

OS534E

-23°C to 871°C (-10°F to 1600°F)

 

OS523E

-18°C to 1371°C (0°F to 2500°F)

 

OS524E

538°C to 2482°C (1000°F to 4500°F)

Accuracy (24°C or 75°F Ambient Temperature and at emissivity of 0.95 or greater):

±1% of reading or

3°F whichever is greater

(2% Rdg for temp > 2000°F for OS524E)

Field of Vision:

OS532E, OS530LE

10:1

 

OS533E, OS530HRE

20:1

 

OS534E

30:1

 

OS53x-CF

0.15"@6"

 

OS523E-1

30:1

 

OS523E-2

60:1

 

OS523E-3

68:1

 

OS524E, OS534E-LR,

110:1

 

OS523E-LR

Repeatability:

± (1% rdg + 1 digit)

 

Resolution:

1°C or 1°F (0.1°C or °F for OS530HRE)

Response Time:

100 msec

 

Spectral Response:

8 to 14 microns (2 to 2.5 microns, OS524)

Thermocouple Input:

Type K, -18 to 871°C (0 to 1600°F)

 

(OS532E, OS533E, OS534E only)

Input Connection:

SMP Connector

 

Thermocouple Display

 

 

Accuracy @ 24°C (75°F)

 

 

Ambient Temperature:

±3°C (±5°F)

 

Thermocouple Display

 

 

Response Time:

2 seconds

 

Operating Ambient

0°C to 50°C

 

Temperature:

(32°F to 122°F)

 

Operating Relative

 

 

Humidity:

95% or less without condensation

Display:

Backlit LCD dual display

 

Keypad:

4 position, tactile feed-back membrane switch

7-1

Image 60
Contents User’sGuide Servicing North America Unpacking Instructions Accessories Page Unpacking Instructions Chapter General Description Chapter Maintenance General Description Introduction OS530LE OS530HRE OS530LE-CF OS533E-CF OS534E-CF Features OS523E OS524E Parts of the Thermometer Front of the Thermometer LCDKey General Description General Description Ac Power Operation Installing the BatteriesOperating the Thermometer Field of View PositionsField of View OS534E, OS523E-1 Field of View OS53xE-CF Field of View OS523E-3 Measurement Techniques 11 Recorder Hookup Real Time Mode Active Operation 12. General Operational Block DiagramReal Time Modes 13. Visual Function Flow Chart Adjusting Emissivity Using the Lock FunctionUsing the Trigger Function Using the Distance Function Must be hard, flat, and reflective to ultrasonic pulseDS = 17. Field of View of Distance Meter HH-DM Calculating Temperature Values Laser Sighting StatusChanging the Temperature from F to C or vice Versa Turning on the Display BacklightingThermocouple Input OS532E, OS533E, OS534E Using the Alarm Functions To set the low alarm value OS533E, OS534E, OS523E, OS524E Use the Ambient Target Temperature Compensation Atcf PC User Application, OS530 Series Operation19. Main Menu Settings Menus 10. Settings Menu21. Typical Temperature Data File PC Interface Commands AsciiString Description Storing Temperature Data on Command OS534E, OS523E, OS524E Logging Temperature Data in Real Time OS523E, OS524E,OS534E Erasing the Temperature Data from Memory Recall Mode Passive Operation ModeRecall Modes Reviewing the Last Parameters Laser Sighting Description Right Side ViewTwo Laser Configurations Operating the Laser SightingLines of Sight of the Laser Sighting and Thermometer Installing and Operating the Sighting Scope Sighting ScopeInstalling the Sighting Scope Safety Warning Maintenance Replacing the BatteriesCleaning the Lens Calibrating the ThermometerServicing the Laser Sighting Troubleshooting Guide Problem SolutionTroubleshooting Guide Problem Troubleshooting Guide Specifications are for all models except where noted SpecificationsGND With Built-in Distance Module Laser Sighting Distance Measuring Built-in-DM Specifications Glossary of Key Strokes Glossary of Key Strokes Thermal Radiation Appendix How Infrared Thermometry Works aBlackbody Calculating Temperature Optics Field of View Metals Appendix Emissivity ValuesNonmetals Appendix Determining an Unknown Emissivity Figure C-1. Determining Emissivity Figure C-2. Determining Emissivity with a Drilled Hole Appendix Determining an Unknown Emissivity Index Field of View Diagrams Recall WARRANTY/DISCLAIMER Temperature PH/CONDUCTIVITY