Omega Engineering manual Accuracy, Pneumatic Operation Not GP901-05 version

Page 4

Network type

Orbit - RS485 2-wire multi-drop (up to 31 per channel)

Network speed

1.5Mbaud, 187.5k Baud or 9600 Baud (do not use 9600

 

Baud for New Applications as this will be phased out)

Network Control Character

Break Character - low (space) of >90 us (187.5k Baud) or

 

>1.2mS (9600 Baud)

Command Character Format

1 start, 8 data, 1 odd parity, 1 stop bit

Accuracy

The error quoted assumes periodic re-setting against a setting master in accordance with normal gauging practice (not less than once per day). As the Digital Probe may be used anywhere on its stroke the setting master itself must be of known dimensions, but does not need to be precisely on its nominal value. The error quoted for the Digital Probe therefore includes a factor dependant on the distance between the setting master and the component being gauged.

Where D = difference in mm between setting master and component.

Error at calibration temperature: -

 

0.05 mm 6 0.2% x D

0.5 mm Probes i.e. GP901-05-S

6

1 mm Probes i.e. GP901-1-S

6

0.1 mm 6

0.1% x D

2 mm Probes i.e. GP901-2-S, GP901-2-P

6

0.1 mm 6

0.1% x D

5 mm Probes i.e. GP901-5-S, GP901-5-P

6

0.1 mm 6

0.2% x D

10 mm Probes i.e. GP901-10-S, GP901-10-P

6

0.1 mm 6

0.2% x D

20 mm Probes i.e. GP901-20-S, GP901-20-P

6

0.1 mm 6

0.2% x D

Gauge Head Specifications

 

 

 

Specifications Common to All Gauge Head Types

 

 

 

Body Material

Stainless Steel 400 Series

Probe tips

various types available

Non-repeat

<0.15 mm

 

Max. Angular Rotation of Tip

2°

 

 

Gaiter Material

Viton®

 

Max. Probe tip tightening Torque

0.22 Nm – M2.5 Thread

(when changing tip please take care not to damage probe) 0.12 Nm – M2 Thread

Storage Temperature Range

-20°C to +100°C

Operating Temperature

+5°C to +80°C

Environmental Protection (see mounting notes)

IP 65 (with gaiter)

Cable Standard (not all types)

2m screened, PUR, insulation

Option (not all types)

Standard cable with stainless steel or plastic braid

Pneumatic Operation (Not GP901-05 version)

Minimum operating pressure

With Gaiter

0.4

(bar)

Without Gaiter

0.3

Maximum operating pressure

With Gaiter

1.0

(bar)

Without Gaiter

2.0

Air leak rate ml/min. at 1.0 bar

With Gaiter

<10

 

Without Gaiter

<150



GP901 M-3343 05/08

Image 4
Contents User’s Guide For immediate technical or application assistance Digital Probe Specifications IntroductionDigital Probe Data GP901-05 GP901-1Accuracy Pneumatic Operation Not GP901-05 versionGauge Head Application Information TipPneumatic Gage Heads Environmental ConsiderationsCable ConnectionsPIE and T-CON Installation This Page Intentionally Left Blank This Page Intentionally Left Blank This Page Intentionally Left Blank WARRANTY/DISCLAIMER Temperature

GP901 specifications

The Omega Engineering GP901 is a notable addition to the realm of industrial instrumentation, specifically designed for high-performance applications in diverse settings. Engineered for accuracy and reliability, the GP901 is particularly well-suited for temperature measurement, boasting a robust set of features that cater to both novice and expert users in various industries.

At the core of the GP901's functionality is its advanced thermocouple technology, which provides highly accurate temperature readings across a wide range of applications. With the ability to measure temperatures from -200°C to 1372°C, the GP901 easily accommodates a variety of processes, from cryogenic environments to high-temperature industrial operations. This versatility makes the device a staple in sectors such as chemical processing, food and beverage production, and HVAC systems.

One of the standout features of the GP901 is its user-friendly interface. The device is equipped with a clear LCD display that ensures real-time monitoring of temperature readings, facilitating quick adjustments when required. Its intuitive design allows users to navigate through settings seamlessly, making configuration and calibration straightforward. This ease of use minimizes downtime, thereby enhancing operational efficiency.

Additionally, the GP901 includes robust connectivity options, allowing it to integrate into existing systems effortlessly. With compatibility for various sensors and communications protocols, including RS-232 and USB, the GP901 is capable of transmitting data to central control systems for further analysis and record-keeping. This capability is essential for industries that require stringent compliance with regulatory standards.

Durability is another hallmark of the GP901, as it features a rugged housing that is resistant to harsh environmental conditions. This makes the unit ideal for outdoor installations or areas with exposure to moisture, dust, or extreme temperatures. The device's design takes into consideration not only performance but also longevity, ensuring that it can withstand the rigors of industrial use.

In conclusion, the Omega Engineering GP901 stands out for its exceptional performance characteristics, user-friendly functionality, versatility in temperature measurement, and durable design. Its commitment to precision and reliability makes it an indispensable tool for professionals across various fields, solidifying its reputation as a leading choice in temperature measurement technology.