TROUBLESHOOTING
Problem | Probable cause | Corrective action | |
Nothing appears in | Batteries are drained. | Replace all batteries with new | |
the display, after I |
| ones. Alkaline batteries | |
press the START |
| recommended. | |
button. |
|
| |
Battery terminals are not in | Reinstall the batteries with | ||
| |||
| the correct position. | negative and positive termi- | |
|
| nals matching those indicated | |
|
| in the battery compartment. | |
The cuff does not | Battery voltage is too low. | Replace all batteries with new | |
inflate. | Low battery symbol blinks. | ones. Alkaline batteries rec- | |
| [If the batteries are drained | ommended. | |
| completely, the mark does |
| |
| not appear.] |
| |
The unit does not | The cuff is not fastened | Fasten the cuff correctly. | |
properly. |
| ||
measure. Readings |
| ||
are too high or too | You moved your arm or | Make sure you remain very | |
low. | |||
body during the | still and quiet during the | ||
| |||
| measurement. | measurement. | |
| The cuff position is not cor- | Sit comfortably and still. | |
| rect. | Make sure the cuff is at the | |
|
| same level as your heart. | |
| You are using the wrong | See Pg. | |
| size cuff. | Correct Cuff." | |
| Taking too many readings | Relax five minutes before each | |
| on the same arm in a short | measurement. | |
| period of time. |
| |
The value is differ- | The healthcare professional | See Pg. | |
ent from that | may be using a different | Why Your Readings Are | |
measured at a clinic | sized cuff. | Different Between Home and | |
or doctor’s office. |
| the Doctor’s Office." | |
|
|
| |
| Your measurements may be | See Pg. | |
| elevated by white coat | Blood Pressure at Home." | |
| hypertension |
| |
|
|
|
NOTE: If the actions described above do not solve the problem, call
LifeSource Health Line |