MEASURING YOUR BLOOD PRESSURE
2 | Plug the air hose | Air hose socket |
| ||
| connector into the main |
|
| unit. |
|
• Plug the air hose connector firmly into the air hose
socket on the main unit, as Air hose connector illustrated.
3Place the cuff at the height of the heart.
4Press the START/STOP switch (the blue switch).
*If the pressurization value exceeds 280 mmHg, if pressurization feels abnormal, or if you want to stop the measurement process, press the START/STOP switch again. The cuff deflates, and the power turns off.
1 All digits on the display blink for approximately one second.
• All digits displayed is an initial display, indicating that the monitor is functioning normal.
The mark will also be displayed but this does not indicate that the battery needs to be replaced.
2* is displayed for approximately one second.
3 The cuff is pressurized automatically.
MEASURING YOUR BLOOD PRESSURE
4 is displayed soon after pressurization is complete, and measurement begins automatically.
*If the unit judges that pressurization is insufficient, it automatically repressurizes.
5 blinks each time a pulse is detected, and the pressure value is displayed on the diastolic blood pressure display.
6 Once measurement is complete, the cuff deflates and the measurement results (systolic/diastolic blood pressure values, pulse) are displayed.
* Systolic/diastolic blood pressure values (two seconds) and pulse (one second) are displayed alternately.
*Provided there is no error in the measurement results, they are stored automatically.
The Body Movement Indicator is displayed on the measurement results display if you move your hand or arm during measurement, causing a substantial pressure change to be detected.
•If the Body Movement Indicator is displayed, measure your blood pressure again.
If an irregular heartbeat is detected during measurement, the Irregular Heartbeat (IHB) Indicator is displayed after measurement is completed.
An irregular heart beat is defined as a heart beat that varies by ±25% from the middle of 5 heart beat intervals during the blood pressure measurement.
• Measurement may not be correct if your heartbeat fluctuates greatly during measurement. If the Irregular Heartbeat (IHB) Indicator is displayed, measure your blood pressure again.
If the Irregular Heartbeat (IHB) Indicator appears Caution frequently, you are advised to consult a doctor about
your health condition.