SECTION 4—TYPICAL PRODUCT PARAMETERS

SECTION 4—TYPICAL PRODUCT PARAMETERS

Electrical Requirements:

120 VAC ± 10% (132 VAC/108 VAC), 60

 

Hz

 

 

Rated Current Input:

Platinum 5 and XL: 4.3 A

 

 

 

Platinum 10: 4.7 A

 

 

Sound Level:

Platinum 5 and XL: 50 dB Average

 

 

 

Platinum 10: 58 dB Average

 

 

Altitude:

Platinum 5 and XL - Up to 6,000 ft (1828

 

meters) above sea level without degradation

 

of concentration levels. From 6,000 ft (1828

 

meters) to 13,129 ft (4000 meters) below

 

90% efficiency

 

 

 

Platinum 10 - Up to 4,000 ft (1230 meters)

 

above sea level without degradation of con-

 

centration levels. From 4,000 ft (1230

 

meters) to 13,129 ft (4000 meters) below

 

90% efficiency

 

 

*Oxygen Output

All 5LXO2/5LX models 87% to 95.6% for

Concentration Levels:

flows ranging 0.5 to 5 L/min and all

 

10LXO2/10LX models 87% to 95.6% for

 

flows ranging 2 L to 10 L/min

 

 

Maximum Outlet Pressure:

Platinum 5 and XL: 5 psi ± 0.5 psi (34.5 kPa

 

± 3.45 kPa)

 

 

 

Platinum 10: 9 psi ± 0.5 psi (62.0 kPa ± 3.45

 

kPa)

*NOTE: Stated concentration levels achieved after initial warm‐up period (approximately 30 minutes).

Part No 1118353

13

PlatinumSeries

Image 13
Contents Platinum Series XL, 5 Accessories Warning AccessoriesTable of Contents Register Your Product Special Notes Signal Word MeaningSpecial Notes IRC5LX and IRC10LX Label LocationGeneral Guidelines Radio Frequency Interference Operating InformationGeneral Guidelines Oxygen Outlet Flowmeter FeaturesUnpacking HandlingInspection StorageTypical Product Parameters Flow Range Platinum 5 and XL 0.5 to 5 L/min For Operating Ambient Operating InstructionsIntroduction Select a LocationPolarized Plug Warning Set-UpHumidifier Compartment Attaching the Oxygen Tubing Power SwitchFlowrate FlowrateInitial Startup of the Concentrator Front Panel Indicators Units with SensO2Front Panel Indicators Units without SensO2 Front Panel Indicators Units without SensO2Routine Maintenance MaintenanceCleaning the Cabinet Preventive Maintenance Record Solution Troubleshooting GuideTroubleshooting Guide Symptom Platinum Series Platinum Series Limited Warranty Limited Warranty USA