SECTION
Flow Range: | Platinum 5 and XL - 0.5 to 5 L/min For |
| flowrates less than 1 L/min, we recommend |
| the use of the Invacare Pediatric Flowmeter |
| IRCPF16 Accessory. |
|
|
| Platinum 10 - 2 L/min to 10 L/min |
|
|
Low Flow Alarm: | Platinum 5 - 0 L/min to 0.5 L/min, |
| Platinum 10 - 0 L/min to 1 L/min |
| Rapid Audible Alarm Beeping (No |
| Accessories Connected) |
|
|
Average Power | Platinum 5 and XL: 400W |
Consumption: | Platinum 10: 585W |
|
|
Pressure Relief Mechanism | 35 psi ± 3.5 psi (241 kPa ± 24.1 kPa) |
Operational at: |
|
|
|
Change in maximum recom- | 0.7 L/min |
mended flow when back pres- |
|
sure of 7kPa is applied: |
|
|
|
Filters: | Cabinet (2), Outlet HEPA and Compressor |
| Inlet. |
|
|
Safety System: | Current overload or line surge shutdown. |
| High temperature compressor shutdown. |
| High Pressure Alarm w/compressor shut- |
| down. Low Pressure Alarm w/compressor |
| shutdown. Battery Free Power Loss Alarm. |
| SensO2 Oxygen System (SensO2 Model) |
| Low Flow Alarm. |
|
|
Width: | 18 3/8 in ± 3/8 in (46.7 cm ± 1 cm) |
|
|
Height: | 26 3/8 in ± 3/8 in (67.0 cm ± 1 cm) |
|
|
Depth: | 14 3/8 in ± 3/8 in (36.5 cm ± 1 cm) |
|
|
Weight: | Platinum 5 and XL: 52 lbs ± 2 lbs (23.6 kg ± |
| 1 kg) |
|
|
| Platinum 10: 53 lbs ± 2 lbs (24.0 kg ± 1 kg) |
|
|
Shipping Weight: | Platinum 5 and XL: 58 lbs ± 2 lbs (26.3 kg ± |
| 1 kg) |
|
|
| Platinum 10: 60 lbs ± 2 lbs (27.2 kg ± 1 kg) |
|
|
Platinum™ Series | 14 | Part No 1118353 |