SECTION
Front Panel Indicators - Units with SensO2
NOTE: For this procedure, refer to FIGURE 5.7.
The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate.
OXYGEN PURITY
Auto Shut - Down
Use Backup | RED | |
Call Supplier | ||
| ||
Below Normal | YELLOW | |
Normal O2 | GREEN |
FIGURE 5.7 Front Panel Indicators - Units with SensO2
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 min.) while waiting for the O2 purity to reach maximum.
LABEL |
| O2 PURITY | INDICATOR LIGHTS |
SYMBOL |
|
| (LED) |
|
|
| |
O2 | SYSTEM OKAY | GREEN Indicator Light | |
O2 over 85% (± 2%) |
| ||
| O Between 73% (± 3%) | YELLOW Indicator light | |
| 2 |
| A. YELLOW Solid |
|
|
| |
| to 85% (±2%) | B. YELLOW Flashing Sensor Failure | |
|
|
| Call a qualified technician. Have a |
|
|
| |
|
|
| |
| SYSTEM FAILURE | RED Indicator Light | |
| O2 Below 73% (±3%) | Continuous Audible Alarm | |
|
|
| |
|
|
| down. Call a qualified technician. |
|
|
| Switch to a |
|
|
| gen immediately. |
Part No 1118353 | 21 | Platinum™ Series |