SECTION 7—TROUBLESHOOTING GUIDE

 

 

 

SYMPTOM:

PROBABLE CAUSE:

SOLUTION:

YELLOW or RED

1. Low oxygen purity.*

1. Clean or Replace filters.

LIGHT Illuminated.

 

 

2. Kinked or blocked tub-

2. Inspect for kinks or block-

Alarm:

ing, cannula or humidifier.*

ages. Correct, clean or replace

 

item. Once corrected, turn

CONTINUOUS

 

power OFF for 60 seconds and

On red light only.

 

then turn power back ON.

 

 

 

* Only applies to

3. Flowmeter set too low:

3. Check flowmeter is set to

0.5 L/min (or less) for

1.0 L/min. or more.

5LXO2 and

Platinum 5; 1.0 L/min (or

 

10LXO2 SensO2

less) for Platinum 10.

 

models.

 

 

4. Unit overheating due to

4.a. Remove and clean cabinet

 

blocked air intake.

filters.

 

 

 

 

 

4b. Move concentrator at least

 

 

three inches from walls, drap-

 

 

eries, and furniture.

 

 

 

 

5. Insufficient power at

5. DO NOT use extension

 

outlet.

cords. Move to another elec-

 

 

trical outlet or circuit.

 

 

 

 

6. Internal repairs

6. Call Service Provider.

 

required.

 

 

 

 

GREEN LIGHT

1. Internal repairs

1. Call Service Provider.

with YELLOW

required.

 

Light Flashing.

 

 

 

 

 

Part No 1118353

27

PlatinumSeries

Image 27
Contents Platinum Series XL, 5 Accessories Warning AccessoriesTable of Contents Register Your Product Special Notes Signal Word MeaningSpecial Notes IRC5LX and IRC10LX Label LocationGeneral Guidelines Radio Frequency Interference Operating InformationGeneral Guidelines Oxygen Outlet Flowmeter FeaturesStorage HandlingUnpacking InspectionTypical Product Parameters Flow Range Platinum 5 and XL 0.5 to 5 L/min For Operating Ambient Select a Location InstructionsOperating IntroductionPolarized Plug Warning Set-UpHumidifier Compartment Attaching the Oxygen Tubing Power SwitchFlowrate FlowrateInitial Startup of the Concentrator Front Panel Indicators Units with SensO2Front Panel Indicators Units without SensO2 Front Panel Indicators Units without SensO2Routine Maintenance MaintenanceCleaning the Cabinet Preventive Maintenance Record Solution Troubleshooting GuideTroubleshooting Guide Symptom Platinum Series Platinum Series Limited Warranty Limited Warranty USA