SECTION 7—TROUBLESHOOTING GUIDE
|
|
| |
SYMPTOM: | PROBABLE CAUSE: | SOLUTION: | |
YELLOW or RED | 1. Low oxygen purity.* | 1. Clean or Replace filters. | |
LIGHT Illuminated. |
|
| |
2. Kinked or blocked tub- | 2. Inspect for kinks or block- | ||
Alarm: | ing, cannula or humidifier.* | ages. Correct, clean or replace | |
| item. Once corrected, turn | ||
CONTINUOUS |
| power OFF for 60 seconds and | |
On red light only. |
| then turn power back ON. | |
|
|
| |
* Only applies to | 3. Flowmeter set too low: | 3. Check flowmeter is set to | |
0.5 L/min (or less) for | 1.0 L/min. or more. | ||
5LXO2 and | |||
Platinum 5; 1.0 L/min (or |
| ||
10LXO2 SensO2 | less) for Platinum 10. |
| |
models. |
|
| |
4. Unit overheating due to | 4.a. Remove and clean cabinet | ||
| blocked air intake. | filters. | |
|
|
| |
|
| 4b. Move concentrator at least | |
|
| three inches from walls, drap- | |
|
| eries, and furniture. | |
|
|
| |
| 5. Insufficient power at | 5. DO NOT use extension | |
| outlet. | cords. Move to another elec- | |
|
| trical outlet or circuit. | |
|
|
| |
| 6. Internal repairs | 6. Call Service Provider. | |
| required. |
| |
|
|
| |
GREEN LIGHT | 1. Internal repairs | 1. Call Service Provider. | |
with YELLOW | required. |
| |
Light Flashing. |
|
| |
|
|
|
Part No 1118353 | 27 | Platinum™ Series |