SECTION 6—MAINTENANCE

SECTION 6—MAINTENANCE

WARNING

The Invacare concentrators are specifically designed to minimize routine preventive maintenance. Only qualified personnel should perform preventive maintenance on the concentrator.

Power should be disconnected before beginning preventive maintenance on the Invacare concentrator.

NOTE: At a minimum, preventive maintenance MUST be performed according to the maintenance record guidelines. In places with high dust or soot levels, maintenance may need to be performed more often. Refer to Preventive Maintenance Record on page 25.

Routine Maintenance

NOTE: The following routine maintenance should be performed more frequently. See recommended intervals in each section and preventive maintenance record.

Cleaning the Cabinet Filter

CAUTION

DO NOT operate the concentrator without the filter installed.

NOTE: For this procedure, refer to FIGURE 6.1 on page 24.

NOTE: There are two cabinet filters one located on each side of the cabinet.

1.Remove each filter and clean at least once a week depending on environmental conditions.

NOTE: Environmental conditions that may require more frequent

cleaning of the filters include but are not limited to: high dust, smoking, air pollutants, etc.

Part No 1118353

23

PlatinumSeries

Image 23
Contents Platinum Series XL, 5 Accessories Warning AccessoriesTable of Contents Register Your Product Special Notes Signal Word MeaningSpecial Notes IRC5LX and IRC10LX Label LocationGeneral Guidelines Radio Frequency Interference Operating InformationGeneral Guidelines Oxygen Outlet Flowmeter FeaturesStorage HandlingUnpacking InspectionTypical Product Parameters Flow Range Platinum 5 and XL 0.5 to 5 L/min For Operating Ambient Select a Location InstructionsOperating IntroductionPolarized Plug Warning Set-UpHumidifier Compartment Attaching the Oxygen Tubing Power SwitchFlowrate FlowrateInitial Startup of the Concentrator Front Panel Indicators Units with SensO2Front Panel Indicators Units without SensO2 Front Panel Indicators Units without SensO2Routine Maintenance MaintenanceCleaning the Cabinet Preventive Maintenance Record Solution Troubleshooting GuideTroubleshooting Guide Symptom Platinum Series Platinum Series Limited Warranty Limited Warranty USA