RELIANT SCALE RLS6
Assembly, Installation and Operating Instructions
SAVE THESE INSTRUCTIONS
NOTE: The Reliant Scale CANNOT be used in conjunction with the swivel bar supplied with the Invacare 9805 Lift. The Reliant Swivel Bar (Part No. 1064518) MUST be purchased and installed on the Invacare 9805 lift in order to install the Reliant Scale.
NOTE: Check all parts for shipping damage before using. In case of damage, DO NOT use the equipment. Contact the Dealer/Carrier for further instructions.
SAFETY SUMMARY
GENERAL WARNINGS
DO NOT install or use this equipment without first reading and understanding these instructions. If you are unable to understand the Warnings, Cautions or Instructions, contact a healthcare professional, dealer or technical personnel before attempting to install this equipment - otherwise, injury or damage may occur.
The weight capacity is limited to the lowest rated capacity of any one of the components in use (i.e. Patient Lift, Sling or Scale). The patient's weight MUST NOT exceed the lowest rated capacity of any component.
After ANY adjustments, repair or service and BEFORE use, make sure all attaching hardware is tightened securely - otherwise injury or damage may occur.
Patient and sling MUST be removed from the lift during ALL installation procedures.
This kit contains the following:
DESCRIPTION | QTY |
|
|
Load Cell Assembly | 1 |
Nylon Washers | 2 |
Swivel Bar Pin | 1 |
Locknut | 1 |
Mounting Screw | 1 |
Small Steel Washer | 1 |
2 | |
2 | |
|
|
Reliant Scale | 1 | Part No. 1114832 Rev B |