6.Note the weight display.

7.When the weight display becomes stable press the LOCK button to lock the weight display. This will be indicated by the word "LOCK" appearing in the display window.

NOTE: Should it be necessary to unlock the weight while the patient is still supported by the lift, the UNLOCK button may be pressed. The weight will unlock and the word "LOCK" will disappear from the display window. The weight value will then be updated. Once the weight becomes stable the weight can be locked again by pressing the LOCK button.

NOTE: Stable being defined as the weight fluctuating two tenths (2/10) of a pound. For example, a patient weighing one hundred (100) lbs the scale will fluctuate between 99.8 and

100.2until the LOCK key is pressed. Fluctuation of the weight displayed is normal as noted above. Press the LOCK button to lock the weight.

NOTE: The UNITS button may be pressed to toggle between units of pounds and kilograms. This is indicated by lb or kg appearing in the display window.

8. The lift may now be lowered and the sling removed from the patient.

NOTES:

The patient's weight will continue to be seen in the display window.

The display will turn off automatically after a two minute period of non-use [no changes in weight exceeding five (5) pounds (2 kilograms)]. You can NOT adjust the time delay for automatic shut off.

After the display has turned off , the weight may be recalled by pressing the ON/OFF button.

The unit can be turned off by pressing the ON/OFF button a second time.

BATTERY REPLACEMENT (FIGURE 4)

NOTE: The scale is powered by a nine (9) volt alkaline battery that should provide approximately 1500 readings before needing replacement.

When battery replacement is needed, LO BAT will appear on the display. Perform the following:

1.Slide the battery door open in the direction of the arrow.

2.Remove existing battery.

3.Install NEW battery.

4.Reinstall the battery door.

Assembled Scale

9V Battery

Battery

Door

Arrow

FIGURE 4 - BATTERY REPLACEMENT

Reliant Scale

7

Part No. 1114832 Rev B

Image 7
Contents Description QTY Safety SummaryReliant Scale RLS6 General WarningsDescription SpecificationsRemove the Swivel BAR Figure Lift Detail a Installing the Reliant Scale FigureKeypad Functions Figure Operating the ScaleReliant Scale Battery Replacement Battery Replacement FigureCalibration of the Reliant Scale Display Codes TroubleshootingLimited Warranty Description Quantité Résumé DE SécuritéBalance Reliant RLS6 Avertissements GénérauxSpécifications Enlever LA Barre À Rotule FigureEnlever LA Barre À Rotule Schéma a Schéma B Installer LA Balance Reliant FigureFonctionnement DE LA Balance Indicateur Touche Affiché Emplacement DéfinitionFonctions DES Touches Figure Avertissement Remplacer LA Pile Remplacer LA Pile FigureCalibrer LA Balance Reliant Codes DépannageGarantie Limitée