QHW digital weighing scales user’s guide
SECTION 11 ERROR CODES
During the initial
If an error message is shown repeat the procedure that caused the message, turning the balance on, calibration or other functions. If the error message still is shown then contact your dealer for further support.
ERROR CODE | DESCRIPTION |
|
| POSSIBLE CAUSES | |
Err 4 | Initial Zero is greater than | Weight on the pan when turning the | |||
| allowed | (4% of | maximum | scale on. | |
| capacity) | when | power | is | Excessive weight on the pan when |
| turned on or when the | zeroing the scale. | |||
| ZERO/ENTER | key | is | Improper calibration of the scale. | |
| pressed, |
|
|
| Damaged load cell. |
|
|
|
|
| Damaged Electronics. |
Err 5 | Keyboard Error. |
|
| Improper operation of the scale. | |
Err 6 | A/D count is not correct | Platform not installed. | |||
| when turning the scale on. |
| Load cell damaged. | ||
|
|
|
|
| Electronics damaged. |
To view the A/D count press the ZERO/ENTER key and the % key at the same time while the scale is performing the initial check at
- 14 -