Quantum Quantum Composers, 9412/9112 A Store/Recall/Misc. Menus, T2Wait, T2*n T2 Period Factor

Page 17

T2:Wait

The output of Channel 2 can be delayed for a number of pulses. This menu is used to set the number of pulses that must occur on the other channels before the output of Channel 2 is en- abled. This is particularly useful when using Channel 2 as a Q-switch trigger, as it allows the laser to thermally stabilize before activating the Q-switch.

This function is active only in Continuous, Burst, and Duty Cycle modes. Channel 2 is delayed at the start of each burst in the Burst mode or at the start of each cycle in the Duty Cycle mode.

T2*n - T2 Period Factor

The period at which Channel 2 (T2) is pulsed can be set to be a multiple of the other channels. This menu item is used to select the multiplier. For example, if the To period ToPer) is set to

.100 sec (10Hz), setting T2*n to 2 will cause the channel 2 period to be .200 sec (5Hz).

Store/Recall/Misc. Menus

The 9412/9412A contains non-volatile memory for storing four (4) (eight (8) for the 9412A) different configurations. The settings of all menu selections are stored and can be quickly re- called.

STORE Configuration

Stores the current menu settings in the specified memory location (one (1) - four

(4) (eight (8) for 9412A).

Note: If you overwrite a particular location, its previous contents are destroyed.

To store a configuration use the Up and Down arrow keys to select the memory location number and press the Left or Right arrow key. You will be prompted to press the key to verify that you want to overwrite the existing stored configuration. If you press any other key, the configuration will not be saved.

17

Image 17
Contents Warranty Parts ListIntroduction Technical SupportCustom Modifications Safety IssuesPage Page Item Button Front Panel Controls and ConnectorsDisplay Control ButtonsBack Panel QC ConnectorsSerial Port Output Voltage AdjustmentAC Input Line Voltage SelectionPage Page Page Item Edit Mode Numeric Input ModeOperating Instructions OverviewMain Menu Mode Pulse ModesGate Gate Select Duty CycleTnDly Tn Delay Width and Delay MenusToPer To Period TnWid Tn Pulse WidthT2Gate T2EnableT2 Menus Store/Recall/Misc. Menus Store ConfigurationT2Wait T2*n T2 Period FactorVolume Recall ConfigurationShot Counter FmarkPage 3 2 Command Protocol Command LanguageComputer Interface Serial RS232 PinoutEcho Enabled Cycles ?crlf 100crlf Communication Protocol Following table summarizes all the commands Command ListCycles 9412/9412A Communication Commands Delays SpecificationsOutputs Replacing the Fuse MaintenancePage Custom Features