Quantum 9412/9112 A, Quantum Composers warranty Duty Cycle, Gate Gate Select

Page 14

Duty Cycle

The Duty Cycle mode is similar to Continuous mode in that a continuous stream of pulses is output. The exception is the outputs are only active for a set number of seconds (DC:On menu), then disabled for another set time period (DC:Off menu) before being re-enabled. This cycle is repeated for a set number of times (CYCLES menu), after which all outputs cease. Setting the number of cycles to zero results in continuous cycling, until the Stop Key is pressed.

The Ext/Gate input may be used as a gate or interlock by setting the GATE menu to Low or High. Pulses will then only be generated while the external trigger input is active. Selecting Low will make the gate active low, while selecting High will make the gate active high.

The Run key must be pressed to start the Duty Cycle mode.

Burst

Burst mode is similar to Continuous, except only a certain number of pulses will be output (#/BURST menu), after which outputs are disabled.

The Ext/Gate input may be used as a gate or interlock by setting the GATE menu to Low or High. Pulses will then only be generated while the external trigger input is active. Selecting Low will make the gate active low, while selecting High will make the gate active high.

The Run key must be pressed to start Burst mode.

GATE - Gate Select

When not in External Trigger mode, the Ext/Gate input on the 9412/9412A can be used as a gate signal. When the gate is enabled, pulses will only be output while the signal into the Ext/Gate input is active -- thus, this signal will act as a safety interlock or external control sig- nal.

The Ext/Gate input may be used as a gate or interlock by setting the GATE menu to Low or High. Pulses will then only be generated while the external trigger input is active. Selecting Low will make the gate active low, while selecting High will make the gate active high. Set- ting the GATE menu to Off will disable the Gate function.

The setting of the GATE menu is used to determine when the signal is active. If the gate menu is set (active) High, pulses will only be generated if the Ext/Gate input signal is greater than 3.5 Volts. If set to (active) Low, outputs will be enabled as long as the Ext/Gate input signal is less than 2.0 Volts.

14

Image 14
Contents Introduction Parts ListWarranty Technical SupportSafety Issues Custom ModificationsPage Page Item Button Display Controls and ConnectorsFront Panel Control ButtonsQC Connectors Back PanelAC Input Output Voltage AdjustmentSerial Port Line Voltage SelectionPage Page Page Operating Instructions Numeric Input ModeItem Edit Mode OverviewMode Pulse Modes Main MenuDuty Cycle Gate Gate SelectToPer To Period Width and Delay MenusTnDly Tn Delay TnWid Tn Pulse WidthT2Gate T2EnableT2 Menus T2Wait Store ConfigurationStore/Recall/Misc. Menus T2*n T2 Period FactorShot Counter Recall ConfigurationVolume FmarkPage 3 2 Computer Interface Command LanguageCommand Protocol Serial RS232 PinoutEcho Enabled Cycles ?crlf 100crlf Communication Protocol Command List Following table summarizes all the commandsCycles 9412/9412A Communication Commands Specifications DelaysOutputs Maintenance Replacing the FusePage Custom Features