HE/001132 | Page 43 of 92 |
| Document No.001132, Revision 2.0 |
Agfa HealthCare | NodeID Livelink : 14140190 |
14 June, 2007 | |
|
|
Attribute Name | Tag | Value |
Execution State Info | (2100, 0030) | Printer dependent. |
Print Priority | (2000, 0020) | HIGH, MED, LOW |
Creation Date | (2100, 0040) | Date of print job creation. |
Creation Time | (2100, 0050) | Time of print job creation |
Printer Name | (2110, 0030) | User defined name identifying the printer |
Originator | (2100, 0070) | DICOM AE title that issued the print operation |
2.2.2.3.1.5SOP Specific Conformance Presentation LUT SOP Class
(1.2.840.10008.5.1.1.23)
The Presentation LUT Information Object is an abstraction of a Presentation LUT. The objective of the Presentation LUT is to realize image display tailored for specific modalities, applications, and user preferences. It is used to prepare image pixel data for display on devices that conform to the Grayscale Standard Display Function.
The output of the Presentation LUT is Presentation Values
A
Table 2.2-35: Supported Attributes for Presentation LUT
Tag | Name | Supported |
| Default |
(2050,0020) | Presentation LUT Shape | IDENTITY | (Note) | IDENTITY |
Note:
‘Presentation LUT Sequence’ is not supported.
"Workstation" (SCU) will process the
Table 2.2-36: N-CREATE confirmation and response Status codes
Code | Status | Meaning |
0000 | Success | Presentation LUT successfully created |
B605 | Warning | Requested Min or Max Density outside of printer’s operating range. The printer will use its |
|
| respective minimum or maximum value instead. |