2.4Configuration
2.4.1AE Title/ Presentation Address Mapping
2.4.1.1Local AE Titles
"Workstation" uses the AE Titles and TCP/IP Ports configured by means of the "Workstation Service & Configuration Tool". The Field Service Engineer can configure the TCP Port via the "Workstation Service & Configuration Tool". No Default AE Titles are provided. The AE Titles must be configured during installation. The local AE Title used by each individual application can be configured independently of the AE Title used by other local applications.
Table 2.4-1: AE Title Configuration Table
Application Entity | Default AE Title | Default TCP/IP Port |
Storage | No Default | 104 |
Hardcopy | No Default | Not Applicable |
RIS | No Default | Not Applicable |
MPPS | No Default | Not Applicable |
2.4.1.2Remote AE Title/ Presentation Address Mapping
The AE title, host names and port numbers of remote applications are configured using the "Workstation Service & Configuration Tool".
2.4.1.2.1Storage
The "Workstation Service & Configuration Tool" must be used to set the AE Titles,
Storage Commit can be enabled or disabled.
The Archive Connection can be tested by means of a DICOM Ping
2.4.1.2.1.1Archive options
When sending images to an archive, this can be done by means of the following SOP classes: CR, DX and MG for presentation or DX and MG for processing. The latter is a RAW image, without any annotations or measurements burned in.
Image pixels can be sent in the requested output type (configurable) depending on the used SOP class and whether GSPS is supported or not6.
CR image, WITHOUT GSPS:
-8 bit OD
-8 bit OD Gamma correction
-8 bit P-Value
6anonymous archiving (i.e. without patient demographics) is not supported