Specifications
Model: | BP 3560/BP 3550 | |
Operation principle: | Oscillometric method | |
Display: | Liquid Crystal Display | |
Measuring range: | 0 – 320 mmHg | (cuff pressure) |
Blood pressure | 30 mmHg | (minimum diastolic value) |
measurement: | 250 mmHg | (maximum systolic value) |
Pulse rate: | 30 – 150 / min |
|
Laboratory accuracy: | ± 3 mmHg | (cuff pressure) |
| ± 5 % of reading (pulse rate) | |
Inflation: | Automatic, soft, adapted to blood pressure | |
Power supply: | 2 batteries, type LR 03 (AAA) | |
| (e.g. Duracell) |
|
Operating temperature: | + 10 °C to + 40 °C | |
Storage temperature: | – 20 °C to + 60 °C | |
Humidity: | Up to 85% r.h. |
|
Cuff: | Suitable for wrist circumference of | |
| 13.5 to 20 cm |
|
If device is not used within specified temperature and humidity ranges the technical accuracy can not be ensured.
Type BF equipment
Subject to change without notice.
13