Braun BP 1650, bp1600 manual Specifications

Page 11

Problem

Reason

Solution

 

 

 

Readings from

Blood pressure is a

none

repeated meas-

fluctuating value. For

 

urements differ

healthy adults, deviations

 

considerably.

of 10 to 20 mmHg are

 

 

possible.

 

 

Cuff is not properly

Ensure that the monitor is fitted as

 

fitted.

follows:

 

 

- on the palm side of your wrist

 

 

- approx. 2 cm away from the

 

 

palm line.

 

Readings were not

For each measurement, hold the

 

consistently taken at

monitor at heart level.

 

heart level.

 

 

Talking, coughing,

While measuring, relax, keep still,

 

laughing, moving etc.

do not move or speak.

 

when measuring will

 

 

influence the reading.

 

 

 

 

Specifications

Model:

BP1650 / BP1600

Operation principle:

Oscillometric method

Display:

Liquid Crystal Display

Measuring range:

0–320 mmHg (cuff pressure)

 

40–199 / min (pulse rate)

Blood pressure

 

measurement:

20 mmHg (minimum diastolic value)

 

300 mmHg (maximum systolic value)

Laboratory accuracy:

± 3 mmHg (cuff pressure)

 

± 5 % of reading (pulse rate)

11

Image 11
Contents VitalScan Plus Braun Infoline SYS What you should know about blood pressure EnglishPage Key rules for accurate blood pressure measurement Product descriptionInserting batteries Taking a measurement Recalling stored readings Memory functionUpper systolic value Lower diastolic value Pulse Time Setting time and date BP 1650 onlyWhat to do if Storage and cleaningProblem Reason Solution CalibrationSpecifications Which are calibrated against auscultatory reference Limited two-year warranty Ce qu’il faut savoir au sujet de la tension artérielle FrançaisPage Description du produit Installation des pilesMesure Rappel des résultats mémorisés Fonction mémoireRéglage de l’heure et de la date BP 1650 seulement Que faire si … Problème Raison SolutionRangement et nettoyage CalibrageProblème Raison Caractéristiques techniques US, CDN Garantie restreinte de deux ans Qué debe saber sobre la presión arterial EspañolPage Inserción de las pilas Descripción del productoMedición Recuperar los datos de la memoria Función de memoriaValor máximo sistólica Valor mínimo diastólica Pulso Selección de fecha y hora solo BP Qué hacer si … Conservación y limpiezaCalibración Solución Problema RazónEspecificaciones Atomo No , Par que Ind. Naucalpan Garantía de dos años Example Exemple Ejemplo Page Mesure Medición Measuring