Setup - Flat Neonatal Panel

This section explains how to prepare a baby for a phototherapy treatment using the flat neonatal fiber optic light panel.

The fiber optic panel must NOT be covered by anything except with the disposable cover provided. Any other type of cover could cause a reduction in light intensity. The setup instructions must be followed exactly.

This panel is ideal for preemie or underweight infants; it may also be used on full-term infants.

1.The protective cover for the neonatal panel is a T-vest. Slide the vertical section of the T, with the light facing the sheer side of the cover, onto the panel.

hook and loop tabs

Neonatal panel with t-vest

2.

Lay the covered panel on a flat surface. Be sure the light

 

emitting side is facing up.

3.

Position the baby’s chest or back directly on the panel. The cable

 

connected to the panel should be between the baby’s legs.

4.

Secure the T-vest to the baby by first wrapping the side without

 

the tape or hook and loop tab around the baby’s midsection.

 

Then, wrap the side with the tape or hook and loop tab over the

infant. If using tape, peel off the protective cover on the tab, and

 

secure it. Be sure the vest is snug.

Positioning the baby on the

 

 

panel

Chapter 3 Setup

Image 13
Contents BiliTx Warranty Table of Contents BiliTx Parents’ Manual Overview What is the BiliTx Phototherapy System?Intended Use Components of the BiliTx System Fiber Optic Light Panel ConfigurationBiliTx BiliTx Parents’ Manual Explanation SymbolsBiliTx Parents’ Manual Setting Up the BiliTx System SetupSetup Wrap-Around Panel Setup Flat Neonatal Panel Using the In-Use Carrying Case Cleaning the Illuminator Device and Fiber Optic Panel CleaningCustomer Service Information Problem Reason/Action TroubleshootingBiliTx Parents’ Manual Environmental SpecificationsPhysical Audible Noise Light SourceStandards Compliance Electrical RequirementsCispr Appendix a EMC InformationImmunity Test IEC 60601 Test Compliance Level Immunity Test IEC 60601 Test Compliance Separation Distance According to Frequency of Transmitter BiliTx Warranty