Light Source

Irradiance Level

Standard Panel-Light:

30µW/cm2/nm

Neonatal Panel-Light:

55µW/cm2/nm

Intensity Ratio

Standard Panel and Neonatal Panel : > .4 (minimum to maximum)

Audible Noise

< 60 dB(A). Measured in accordance with IEC 60601-2-50.

Standards Compliance

This device is designed to conform to the following standards:

IEC 60601-1 General Requirements for Safety of Medical Electrical Equipment

IEC 60601-2-50 Requirements for the Safety of Infant Phototherapy Equipment

Electromagnetic Compatibility: EN 60601-1-2, 2nd edition.

Electrical Requirements

AC Power

100-240 VAC, 50/60 Hz, 1.0 A

Type of Protection Against Electrical Shock

Class I Equipment

 

 

Degree of Protection Against Electrical Shock

Type BF Applied Part

 

 

Degree of Protection Against Ingress of Water

Ordinary Protection, IPX0

Mode of Operation

Continuous

Disposal

Dispose of this device in accordance with local regulations.

BiliTx Parents’ Manual

Image 20
Contents BiliTx Warranty Table of Contents BiliTx Parents’ Manual Intended Use What is the BiliTx Phototherapy System?Overview Fiber Optic Light Panel Configuration Components of the BiliTx SystemBiliTx BiliTx Parents’ Manual Symbols ExplanationBiliTx Parents’ Manual Setup Setting Up the BiliTx SystemSetup Wrap-Around Panel Setup Flat Neonatal Panel Using the In-Use Carrying Case Cleaning Cleaning the Illuminator Device and Fiber Optic PanelCustomer Service Information Troubleshooting Problem Reason/ActionBiliTx Parents’ Manual Physical SpecificationsEnvironmental Light Source Audible NoiseStandards Compliance Electrical RequirementsAppendix a EMC Information CisprImmunity Test IEC 60601 Test Compliance Level Immunity Test IEC 60601 Test Compliance Separation Distance According to Frequency of Transmitter BiliTx Warranty