Fluke 561s user manual Field of View, Emissivity

Page 11

Operating the Thermometer

 

132 mm @

75 mm @

1500 mm

S

900 mm

 

38 mm @

300 mm

5.3 " @

3 " @ 60 "

1.5 " @ 36 "

12 "

D

D:S=12:1

efh005f.eps

Figure 5. Distance and Spot Size

Field of View

Make sure that the target is larger than the spot size. The smaller the target, the closer you should be to it. See Figure 6.

YesNo

efh004f.eps

Figure 6. Field of View

Emissivity

Emissivity describes the energy-emitting characteristics of materials. Most organic materials and painted or oxidized surfaces have an emissivity of about 0.95.

If possible, to compensate for inaccurate readings that may result from measuring shiny metal surfaces, cover the surface to be measured with masking tape or flat black paint

(<148 °C/300 °F) and use the high emissivity setting. Allow time for the tape or paint to reach the same temperature as the surface beneath it. Measure the temperature of the tape or painted surface.

If you cannot paint or use tape, then you can improve the accuracy of your measurements with the emissivity selector. Even with the emissivity selector, it can be difficult to get a completely accurate infrared measurement of a target with a shiny or metallic surface. Experimentation, using the probe to determine benchmark temperatures, and experience will help you choose the best setting for specific measurements.

The Thermometer has three emissivity settings: low (0.3), medium (0.7), and high (0.95). Refer to Table 2. The reference to emissivity settings in the table are suggestions for typical situations. Your particular situation may differ.

7

Image 11
Contents 561 Limited Warranty and Limitation of Liability Table of Contents 561 Contacting Fluke IntroductionXWWarning Safety InformationInfrared Thermometer FeaturesDisplay When the battery is low, B appears on the displayButtons and Connector Buttons and ConnectorDistance and Spot Size How the Thermometer WorksOperating the Thermometer Locating a Hot or Cold SpotEmissivity Field of ViewSwitching Between C and F XWWarning Using the Contact Temperature ProbeCleaning the Housing MaintenanceChanging the Battery Cleaning the LensCE Certification TroubleshootingSpecifications Electrical