RayTek FR, FA Installing of Multiple Sensors in a Network, Wiring, Addressing

Page 32

Installation

5.8 Installing of Multiple Sensors in a Network

5.8.1 Wiring

For an installation of two or more sensors in a network, each sensor cable is wired to its own terminal block. The RS485 terminals on each terminal block are wired in parallel.

The following figure illustrates the wiring of sensors in a 4‐wire multidrop installation. A network as a 2‐wire multidrop installation is to realize according to Figure 19, p. 23.

 

from additional terminal block

 

of another sensor

RS232/485

RxB RxA TxB TxA

Interface Converter

 

XXX485CVT…

 

 

from RxB

 

from RxA

 

from TxB

 

from TxA

 

Ground

 

+24 VDC

Elektronikbox

Terminal Block

Electronics Housing

9 VDC or ...

24 VDC power supply

Figure 21: 4‐Wire Multidrop Wiring in a Network

5.8.2 Addressing

The addressing of a sensor can be done by means of the Multidrop Software (Menu <Sensor Setup>) that came with your sensor. As alternative the specific interface commands of the sensor can be used in conjunction with a standard terminal program (e.g. Windows HyperTerminal), see section 9.5 Command List, p. 42.

If you are installing two or more sensors in a multi‐drop configuration, please be aware of the following:

Each sensor must have a unique address.

Each sensor must be set to the same baud rate.

24

Marathon Series FA/FR

Image 32
Contents Marathon FA/FR Series Page Fluke Service Center ContactsWorldwide Headquarters European Headquarters France United KingdomWarranty Table of Contents Milliamp Output Relay Outputs Trigger Fiber Calibration Page Marathon Series FA/FR Safety InstructionsSafety Instructions Product Description Product DescriptionTheory of Operation for 2‐Color Sensors ModelsLow or Changing Emissivities Partially Obscured TargetsTargets Smaller Than Field of View Technical Data Technical DataMeasurement Specifications FA ModelsGeneral Specifications Electrical Specifications Dimensions Dimensions of Optical HeadFA ModelsAdjustable Mounting Bracket for Optical Head Optical Specifications Standard FocusClose Focus Spot Size Charts for FA models Close FocusStandard Focus Spot Size Charts for FR models Close Focus Spot Size Charts for FR models Scope of DeliverySensor Location Sensor LocationSensor Placement 1‐Color Mode Sensor Placement 2‐Color ModeSensor Placement in 2‐Color Mode Viewing AnglesAcceptable Sensor Viewing Angles Mounting the Sensor InstallationInstallation Aiming Fiber Optic CableInstalling the Electronics Housing Configuring the Sensor CableRemoving the Terminal Block RS232/485 Interface Converter Power SupplyConnecting to a PC Multidrop Installation 4‐Wire Multidrop Installation 2‐WireConnecting to Terminal Block Wiring for 2‐Wire Sensor SetupAddressing Installing of Multiple Sensors in a NetworkWiring Control Panel OperationOperation Mode Selector Button Sequence FA Models Operation ModesSlope 2‐Color Temperature DisplayEmissivity 1‐Color Averaging AVG 4 2C/1C SwitchPeak Hold PKH Valley Hold VAL Overview to Hold FunctionsHold Functions SetpointsDeadband Ambient Background Temperature Compensation FA Models Inputs and OutputsMilliamp Output Relay OutputsFactory Defaults Factory DefaultsOptions Cooling Platform for Electronics HousingOptions Accessories AccessoriesAccessories selection OverviewAir Purge Collar Protection TubeFitting System Flexible Fitting SystemOrder number Model Available RS232/485 Interface ConvertersIndustrial Power Supply Dimension of Industrial Power SupplyCommand Structure Programming GuideProgramming Guide Remote versus Manual ConsiderationsTransfer Modes Poll ModeBurst Mode Response Time in Setup Mode = + n ⋅15000Command List Command ListXUFR1A Command Examples Command ExamplesMaintenance MaintenanceTroubleshooting Minor Problems TroubleshootingFail‐Safe Operation Error Codes in 1‐Color Mode FA modelsError Codes in 2‐Color Mode FR models Current Output Values in accordance to an Error Examples of failsafe conditionsCleaning the Lens Removing the Fiber Optic Cable from the Optical Head Replacing the Fiber Optic CableRemoving the Fiber Optic Cable Put cable out Mounting the Fiber Optic Cable Attaching the Fiber Optic Cable to the Optical HeadFiber Calibration Dialog for the Calibration of the Fiber Cable Appendix AppendixDetermination of Emissivity Typical Emissivity ValuesTypical Emissivity Values Non‐Metals Marathon Series FA/FR55Typical Emissivity Values Metals How to determine slope? Typical SlopesSignal Reduction FR Models Typical Percentage of Allowed Signal Reduction FR1A ModelsMaximum Error Attenuation Influence on AccuracyMarathon Instrument Calibration