RayTek FR, FA operating instructions Command Examples

Page 52

Programming Guide

9.6 Command Examples

 

 

HOST

 

 

SENSOR

 

 

HOST

 

 

SENSOR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WHERE USED (1)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description

 

Query

 

 

Answer

 

 

Set

 

 

Notification

 

 

P

 

B

S

N

 

Burst string format

001?$

 

 

001!$UTSI

 

 

001$=UTSI

 

 

 

 

 

 

 

 

 

Show list of commands

001?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measured attenuation

 

001?B

 

 

001!B12

 

 

 

 

 

 

 

 

 

 

 

 

Baud rate

 

 

 

 

001!D384

 

 

001D=384

 

 

 

 

 

 

 

 

 

 

Emissivity

 

001?E

 

 

001!E0.95

 

 

001E=0.95

 

 

001#E0.95

 

 

 

 

Average time

 

001?G

 

 

001!G001.2

 

 

001G=001.2

 

001#G001.2

 

 

 

 

Top of mA range

 

001?H

 

 

001!H2000

 

 

001H=2000

 

 

001#H2000

 

 

 

 

 

Sensor internal ambient

 

001?I

 

 

001!I028

 

 

 

 

 

 

 

 

 

 

 

 

Switch panel lock

 

001?J

 

 

001!IJL

 

001J=L

 

 

 

 

 

 

 

 

 

Relay alarm output control

 

 

 

 

001!K0

 

 

001K=0

 

 

 

 

 

 

 

 

 

Bottom of mA range

 

001?L

 

 

001!L1200

 

001L=1200

 

 

 

 

 

 

 

 

Mode – FR series

 

001?M

 

 

001!M1

 

 

001M=1

 

 

001#M1

 

 

 

 

Target temperature, 1-color narrow

 

001?N

 

 

001!N1158

 

 

 

 

 

 

 

 

 

 

 

 

 

Output current

 

 

 

 

001!O10

 

 

001O=10

 

 

 

 

 

 

 

 

Peak hold time

 

001?P

 

 

001!P005.6

 

 

001P=005.6

 

001#P005.6

 

 

 

 

Power

 

001?Q

 

 

001!Q0036.102

 

 

 

 

 

 

 

 

 

 

 

 

Narrow Power

 

001?R

 

 

001!R0002.890

 

 

 

 

 

 

 

 

 

 

 

 

Slope

 

001?S

 

 

001!S0.850

 

 

001S=0.850

 

 

001#S0.850

 

 

 

 

Target temperature, 2-color (FR only)

 

001?T

 

 

001!T1225

 

 

 

 

 

 

 

 

 

 

 

 

Temperature units

 

001?U

 

001!UC

 

 

001U=C

 

 

001#UC

 

 

 

 

Poll/Burst mode

 

 

 

 

001!VP

 

 

001V=P

 

 

 

 

 

 

 

 

 

 

Target temperature, 1-color wide

 

001?W

 

 

001!W1210

 

 

 

 

 

 

 

 

 

 

 

 

Burst string contents

 

001?X$

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Multidrop address

 

001?XA

 

 

001!XA013

 

 

001XA=013

 

 

 

 

 

 

 

 

Low temperature limit

 

001?XB

 

 

001!XB

 

 

 

 

 

 

 

 

 

 

 

 

 

Deadband

 

001?XD

 

 

001!XD12

 

 

001XD=12

 

 

 

 

 

 

 

 

 

Restore factory defaults

 

 

 

 

001!XF

 

 

XF

 

 

001#XF

 

 

 

 

 

 

High temperature limit

 

001?XH

 

 

001!XH1400

 

 

 

 

 

 

 

 

 

 

 

 

 

Sensor initialization

 

001?XI

 

 

001!XI0

 

 

001XI=0

 

 

001#XI

 

 

 

 

Laser

 

001?XL

 

 

001!XL1

 

 

001XL=1

 

 

001#XL1

 

 

 

 

 

Sensor model type

 

001?XM

 

001!XR

 

 

 

 

 

 

 

 

 

 

 

 

 

0-20 mA or 4 - 20 mA analog output

 

001?XO

 

 

001!XO4

 

 

001XO=4

 

 

 

 

 

 

 

 

 

Second setpoint

 

001?XP

 

 

001!XP1234

 

001XP=1234

 

 

 

 

 

 

 

 

 

Sensor revision

 

001?XR

 

 

001!XRF1

 

 

 

 

 

 

 

 

 

 

 

 

 

Setpoint / relay function

 

001?XS

 

 

001!XS1234

 

 

001XS=1234

 

 

 

 

 

 

 

 

 

Trigger

 

001?XT

 

 

001!XT0

 

 

 

 

 

001#XT0

 

 

 

 

 

Identify unit

 

001?XU

 

 

001!XUFR1

 

 

 

 

 

 

 

 

 

 

 

 

 

Sensor serial number

 

001?XV

 

 

001!XVA099901

 

 

 

 

 

 

 

 

 

 

 

 

 

Attenuation to activate relay

 

001?Y

 

 

001!Y95

 

 

001Y=95

 

 

 

 

 

 

 

 

Attenuation for failsafe

 

001?Z

 

 

001!Z99

 

 

001Z=99

 

 

 

 

 

 

 

 

Table 7: Command Examples

(1)P = Poll Mode (Request for a parameters), B = Burst Mode (continuous sending of parameters in the burst string), S = Set (Command for setting a parameters), N = Notification (Acknowledgment for setting a parameter)

The given examples are related to a unit in a network addressed with address 001. Stand‐alone units are requested without having an address information in the command.

44

Marathon Series FA/FR

Image 52
Contents Marathon FA/FR Series Page Fluke Service Center ContactsWorldwide Headquarters European Headquarters France United KingdomWarranty Table of Contents Milliamp Output Relay Outputs Trigger Fiber Calibration Page Safety Instructions Safety InstructionsMarathon Series FA/FR Product Description Product DescriptionTheory of Operation for 2‐Color Sensors ModelsTargets Smaller Than Field of View Partially Obscured TargetsLow or Changing Emissivities Technical Data Technical DataMeasurement Specifications FA ModelsGeneral Specifications Electrical Specifications Dimensions Dimensions of Optical HeadFA ModelsAdjustable Mounting Bracket for Optical Head Optical Specifications Standard FocusClose Focus Spot Size Charts for FA models Close FocusStandard Focus Spot Size Charts for FR models Close Focus Spot Size Charts for FR models Scope of DeliverySensor Location Sensor LocationSensor Placement 1‐Color Mode Sensor Placement 2‐Color ModeSensor Placement in 2‐Color Mode Viewing AnglesAcceptable Sensor Viewing Angles Installation InstallationMounting the Sensor Aiming Fiber Optic CableInstalling the Electronics Housing Configuring the Sensor CableRemoving the Terminal Block Connecting to a PC Power SupplyRS232/485 Interface Converter Multidrop Installation 4‐Wire Multidrop Installation 2‐WireConnecting to Terminal Block Wiring for 2‐Wire Sensor SetupWiring Installing of Multiple Sensors in a NetworkAddressing Operation OperationControl Panel Mode Selector Button Sequence FA Models Operation ModesEmissivity 1‐Color Temperature DisplaySlope 2‐Color Peak Hold PKH 4 2C/1C SwitchAveraging AVG Valley Hold VAL Overview to Hold FunctionsDeadband SetpointsHold Functions Ambient Background Temperature Compensation FA Models Inputs and OutputsMilliamp Output Relay OutputsFactory Defaults Factory DefaultsOptions Cooling Platform for Electronics HousingOptions Accessories AccessoriesAccessories selection OverviewAir Purge Collar Protection TubeFitting System Flexible Fitting SystemOrder number Model Available RS232/485 Interface ConvertersIndustrial Power Supply Dimension of Industrial Power SupplyCommand Structure Programming GuideProgramming Guide Remote versus Manual ConsiderationsBurst Mode Poll ModeTransfer Modes Response Time in Setup Mode = + n ⋅15000Command List Command ListXUFR1A Command Examples Command ExamplesMaintenance MaintenanceTroubleshooting Minor Problems TroubleshootingError Codes in 2‐Color Mode FR models Error Codes in 1‐Color Mode FA modelsFail‐Safe Operation Current Output Values in accordance to an Error Examples of failsafe conditionsCleaning the Lens Removing the Fiber Optic Cable Replacing the Fiber Optic CableRemoving the Fiber Optic Cable from the Optical Head Put cable out Mounting the Fiber Optic Cable Attaching the Fiber Optic Cable to the Optical HeadFiber Calibration Dialog for the Calibration of the Fiber Cable Appendix AppendixDetermination of Emissivity Typical Emissivity ValuesTypical Emissivity Values Metals Marathon Series FA/FR55Typical Emissivity Values Non‐Metals How to determine slope? Typical SlopesSignal Reduction FR Models Typical Percentage of Allowed Signal Reduction FR1A ModelsMaximum Error Attenuation Influence on AccuracyMarathon Instrument Calibration