RayTek operating instructions Technical Data, Measurement Specifications, FA Models, FR Models

Page 12

Technical Data

3 Technical Data

3.1 Measurement Specifications

3.1.1 FA Models

Temperature Range

 

FA1A

475 to 900°C (887°F to 1652°F)

FA1B

800 to 1900°C (1472°F to 3452°F)

FA1C

1200 to 3000°C (2192°F to 5432°F)

FA1G

750 to 1675°C (1382°F to 3047°F)

FA2A

250 to 800°C (482°F to 1472°F)

FA2B

400 to 1700°C (752°F to 3092°F)

Spectral Response

 

FA1

1.0 μm (Si detector)

FA2

1.6 μm (InGaAs detector)

System Accuracy1

 

FA1/FA2

±(0.3% Tmeas + 2°C), Tmeas in °C

FA1G

±3°C

Repeatability

±1°C

Temperature Resolution

 

Current Output

±0.05°C

 

±0.01°C for FA1G

Display and RS485

±1°C

Response Time

10 msec (95%), selectable to 10 sec

Temperature Coefficient

±0.03% full scale change per 1°C change in ambient

 

temperature

Noise Equivalent Temp. (NET)

1°C peak to peak, at target emissivity of 1.00

Emissivity

0.10 to 1.00, in 0.01 increments

Signal Processing

Peak Hold, valley hold, averaging

 

Hold time 0 – 300 sec, in 0.1 sec increments

3.1.2 FR Models

Temperature Range

 

 

FR1A

500

– 1100°C (930°F to 2010°F)

FR1B

700

– 1500°C (1290°F to 2730°F)

FR1C

1000 – 2500°C (1830°F to 4530°F)

Spectral Response

1.0 μm nominal (Si/Si sandwich detector)

1at ambient temperature 23°C ±5°C (73°F ±9°F)

4

Marathon Series FA/FR

Image 12
Contents Marathon FA/FR Series Page Fluke Service Center ContactsWorldwide Headquarters European Headquarters France United KingdomWarranty Table of Contents Milliamp Output Relay Outputs Trigger Fiber Calibration Page Safety Instructions Safety InstructionsMarathon Series FA/FR Product Description Product DescriptionTheory of Operation for 2‐Color Sensors ModelsPartially Obscured Targets Targets Smaller Than Field of ViewLow or Changing Emissivities Technical Data Technical DataMeasurement Specifications FA ModelsGeneral Specifications Electrical Specifications Dimensions Dimensions of Optical HeadFA ModelsAdjustable Mounting Bracket for Optical Head Optical Specifications Standard FocusClose Focus Spot Size Charts for FA models Close FocusStandard Focus Spot Size Charts for FR models Close Focus Spot Size Charts for FR models Scope of DeliverySensor Location Sensor LocationSensor Placement 1‐Color Mode Sensor Placement 2‐Color ModeSensor Placement in 2‐Color Mode Viewing AnglesAcceptable Sensor Viewing Angles Installation InstallationMounting the Sensor Aiming Fiber Optic CableInstalling the Electronics Housing Configuring the Sensor CableRemoving the Terminal Block Power Supply Connecting to a PCRS232/485 Interface Converter Multidrop Installation 4‐Wire Multidrop Installation 2‐WireConnecting to Terminal Block Wiring for 2‐Wire Sensor SetupInstalling of Multiple Sensors in a Network WiringAddressing Operation OperationControl Panel Mode Selector Button Sequence FA Models Operation ModesTemperature Display Emissivity 1‐ColorSlope 2‐Color 4 2C/1C Switch Peak Hold PKHAveraging AVG Valley Hold VAL Overview to Hold FunctionsSetpoints DeadbandHold Functions Ambient Background Temperature Compensation FA Models Inputs and OutputsMilliamp Output Relay OutputsFactory Defaults Factory DefaultsCooling Platform for Electronics Housing OptionsOptions Accessories AccessoriesAccessories selection OverviewAir Purge Collar Protection TubeFitting System Flexible Fitting SystemOrder number Model Available RS232/485 Interface ConvertersIndustrial Power Supply Dimension of Industrial Power SupplyCommand Structure Programming GuideProgramming Guide Remote versus Manual ConsiderationsPoll Mode Burst ModeTransfer Modes Response Time in Setup Mode = + n ⋅15000Command List Command ListXUFR1A Command Examples Command ExamplesMaintenance MaintenanceTroubleshooting Minor Problems TroubleshootingError Codes in 1‐Color Mode FA models Error Codes in 2‐Color Mode FR modelsFail‐Safe Operation Current Output Values in accordance to an Error Examples of failsafe conditionsCleaning the Lens Replacing the Fiber Optic Cable Removing the Fiber Optic CableRemoving the Fiber Optic Cable from the Optical Head Put cable out Mounting the Fiber Optic Cable Attaching the Fiber Optic Cable to the Optical HeadFiber Calibration Dialog for the Calibration of the Fiber Cable Appendix AppendixDetermination of Emissivity Typical Emissivity ValuesMarathon Series FA/FR55 Typical Emissivity Values MetalsTypical Emissivity Values Non‐Metals How to determine slope? Typical SlopesSignal Reduction FR Models Typical Percentage of Allowed Signal Reduction FR1A ModelsMaximum Error Attenuation Influence on AccuracyMarathon Instrument Calibration