Omega Vehicle Security HH508 manual Calibration Procedure

Page 18

HH501(A,B)K / HH501(A,B)J / HH501(A,B)T / HH501(A,B)E

HH508BR/HH508/HH509/HH509R

CALIBRATION PROCEDURE

Note: The following calibration procedure should perform only by qualified technicians who have access to the items as following.

Equipment: The class of calibrator had better 10 times greater than the measured meter.

1.Set the jumper position on J1 position.

2.Input DCV 25.6mV to the T1 after the display is stabilized (forward 4 digits), then press “ ENTER “ key and "01" is displayed.

3.Input DCV 74mV after the display is stabilized, then press “ENTER “ key and "OL" is displayed.

4.Input 0￿C (K-type) after the display is stabilized, then press “ENTER “ key.

5.Don't push any key and the meter will turn off automatically after 10 seconds.

6.After adjustment set the jumper back to J3 position.

7.Turn on the meter then input 0￿C (K-type). The display reading reads 0°C if the calibration procedure is right.

1

Image 18
Contents Omega Omega.comTM Safety Information IntroductionSpecifications EnvironmentalElectrical Hold Mode Power SwitchF Selecting the Temperature Scale Operating InstructionsMIN MAX record Mode 9 T1 T2/T1-T2 for HH501BR/HH501BSREL Relative mode SET mode Relative value set and Hi/Lo Limits value set 12 Hi/Lo Limits mode Operator Maintenance Battery ReplacementRS-232 Operation Using Quick Basic Close #1 Open COM11200,E,7,1,DS,RS for Random AS #1END if RS-232 Transmit and Receive Comammand Summary Output Data Formats Page Communication Calibration Procedure Warranty / Disclaimer 3203/0799