Temperature Scale: Celsius or Fahrenheit user-selectable.

Measurement Range:

Thermocouple

Range

K-TYPE(0.1°C)

-200°C to 1372°C, -328°F to 1999°F

J-TYPE(0.1°C)

-210°C to 1200°C, -346°F to 1999°F

T-TYPE(0.1°C)

-200°C to 400°C, -328°F to 752°F

E-TYPE(0.1°C)

-220°C to 1000°C, -364°F to 1832°F

S-TYPE(1°C)

0°C to 1767°C, 32°F to 1999°F

Auto range: 0.1°C/1°C, 0.1°F/1°F.

Accuracy: Accuracy is specified for operating temperatures over the range of 18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.

 

±(0.1% rdg + 1°C) range -60°C to 1372°C

 

±(0.1% rdg + 2°C) range -60°C to -220°C

 

±(0.1% rdg + 2°F) range -76°F to 1999°F

 

±(0.1% rdg + 4°F) range -76°F to -364°F

S-TYPE

±(0.0% rdg + 2°C) on 0°C to 50°C

 

±(0.1% rdg + 1°C) on 50°C to 1767°C

 

±(0.0% rdg + 4°F) on 32°F to 122°F

 

±(0.1% rdg + 2°F) on 122°F to 1999°F

3

Image 5
Contents Omega Omega.comTM Introduction Safety InformationEnvironmental SpecificationsElectrical F Selecting the Temperature Scale Power SwitchHold Mode Operating Instructions9 T1 T2/T1-T2 for HH501BR/HH501BS MIN MAX record ModeREL Relative mode SET mode Relative value set and Hi/Lo Limits value set 12 Hi/Lo Limits mode Battery Replacement Operator MaintenanceClose #1 Open COM11200,E,7,1,DS,RS for Random AS #1 RS-232 Operation Using Quick BasicEND if RS-232 Transmit and Receive Comammand Summary Output Data Formats Page Communication Calibration Procedure Warranty / Disclaimer 3203/0799