OUTPUT DATA FORMATS

The data format consists of 32 bytes.

byte

0

0

0

0

0

0

0

0

 

0

1

1

1

1

1

1

1

1

2

3

4

5

6

7

8

 

9

0

1

2

3

4

5

6

 

 

 

 

 

 

 

S H H H H H H A

S H H H H H H A

 

 

T1

present

status

 

 

 

 

 

T2

present

status

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

1

1

2

2

2

2

2

2

2

2

2

2

3

3

3

7

8

9

0

1

2

3

4

5

6

7

8

9

0

1

2

D D D D

D D

A A A A A A A A

CR nl

 

Timer

 

 

Meter status

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

S:Negative sign or positive sign

H:Hexadecimal digits

D:Numeric digits

A:Capital Letter or "-" sign CR: Carriage return character nl: newline character

01

 

 

 

 

T1 polarity

 

 

 

"+"

Positive

 

 

 

 

 

 

09

 

 

 

 

T2 polarity

 

 

 

"-"

Negative

 

 

 

 

 

 

 

02

— MSD

 

 

 

 

T 1 Values=[H07x160+H06x161+H05x162

 

 

 

 

07

— LSD

 

 

 

 

+H 04x163+H03x164+H02x165] / 1000

 

 

 

 

10

— MSD

 

 

 

 

T 2 Values=[H15x160+H14x161+H13x162

15

— LSD

 

 

 

 

+H 12x163+H11x164+H10x165] / 1000

 

 

 

 

08 — Thermocouple Type

 

 

 

"K"

®

K type

 

 

 

 

 

 

"J"

®

J type

 

 

 

 

 

 

 

"T"

®

T type

 

 

 

 

 

 

 

 

"E"

®

E type

 

 

 

 

 

 

 

 

"S"

®

S type

 

 

 

 

13

Image 15
Contents Omega Omega.comTM Introduction Safety InformationEnvironmental SpecificationsElectrical Operating Instructions Power SwitchF Selecting the Temperature Scale Hold Mode9 T1 T2/T1-T2 for HH501BR/HH501BS MIN MAX record ModeREL Relative mode SET mode Relative value set and Hi/Lo Limits value set 12 Hi/Lo Limits mode Battery Replacement Operator MaintenanceClose #1 Open COM11200,E,7,1,DS,RS for Random AS #1 RS-232 Operation Using Quick BasicEND if RS-232 Transmit and Receive Comammand Summary Output Data Formats Page Communication Calibration Procedure Warranty / Disclaimer 3203/0799