2.4 Source Mode Procedure (Calibration Simulation)

1.Turn CL3512A on by turning selector switch to any one of the output selector positions.

2.Press T1 button.

3.Select thermocouple type (J, K, T or E) using type button. Select °C or °F using °C or °F buttons.

4.Plug proper type of thermocouple wire into either one of the two output connectors. (SMP male connector required)

5.Attach the other end of the thermocouple wire to the instrument being tested/calibrated.

6.Move the output range switch to the temperature value close to the desired simulated value.

Note: There are 9 temperature range switch positions arranged from left to right representing low to high temperatures.

The display will read -OL if the output voltage for the meter is higher than the maximum voltage for that particular thermocouple type. For type T this will be in the highest 4 positions, type K in the highest 3 positions, type J in the highest 2 positions, and type E in the highest position only.

7.Adjust course and/or fine adjustment dial until desired temperature is displayed on CL3512A.

8.Instrument under test should read same temperature as CL3512A. If not adjust instrument under test or have recalibrated.

9.Repeat steps 6-8 at different temperatures as needed.

7

Image 10
Contents CL3512A Digital Thermometer & Calibrator Omega.comTM Table of contents Safety Information IntroductionUnpacking Operation Proceedure Drawing of UnitButtons Descriptions of Buttons and SwitchesButton Hold Mode ButtonAdjust Selector SwitchInput Selections Read Mode Proceedure Source Mode Procedure Calibration Simulation Source Mode Diagram Operator Maintenance Battery ReplacementSpecifications ElectricalGeneral CL3152A Calibration Procedure Input Funtion CalibrationSimulated Output Test CL3512A Calibration Procedures Warranty / Disclaimer OMEGA...Of Course