Omega Vehicle Security CL3512A manual Simulated Output Test

Page 16

"Simulated Output" Test

1.Turn the knob "COARSE" right to the end, knob of RANGE turn right to the last two position, press "TYPE" key to select "TYPE-E", if display does not show "OL" then adjust VR4 slowly to make display ="OL" (Note: Take care not to adjust VR4 too far).

13

Image 16
Contents CL3512A Digital Thermometer & Calibrator Omega.comTM Table of contents Safety Information IntroductionUnpacking Operation Proceedure Drawing of UnitDescriptions of Buttons and Switches Button Hold ModeButtons ButtonAdjust Selector SwitchInput Selections Read Mode Proceedure Source Mode Procedure Calibration Simulation Source Mode Diagram Operator Maintenance Battery ReplacementSpecifications ElectricalGeneral CL3152A Calibration Procedure Input Funtion CalibrationSimulated Output Test CL3512A Calibration Procedures Warranty / Disclaimer OMEGA...Of Course