SECTION 6—TROUBLESHOOTING
SYMPTOMS
FAULTS
SOLUTION
Noisy or dry sound from pivots. | Needs lubrication. | Refer to Lubricating the Lift on page 22. |
|
|
|
Electric actuator fails to lift when button is | Check the hand control and actuator con- | |
pressed. | connector loose. | nections. |
| Battery low. | Charge batteries. Refer to Charging the Bat- |
|
| tery on page 13. |
| RED emergency stop button | Rotate RED emergency stop button |
| pressed IN. | CLOCKWISE until it pops out. |
| Battery not connected properly | Reconnect the battery to the control box. |
| to control box. | Refer to Charging the Battery on page 13. |
| The connecting terminals are | Replace the battery pack. Refer to Charging |
| damaged. | the Battery on page 13. |
| Electric actuator in need of ser- | Refer to Adjusting the Base on page 22. |
| vice or load is too high. | Contact Dealer if service is required. |
|
|
|
Unusual noise from actuator. | Actuator is worn or damaged | Replace the actuator or contact Dealer. |
| or spindle is bent. | Refer to Replacing the Electric Actuator on |
|
| page 23. |
|
|
|
Lift arms will not lower from the uppermost | Lift arms require a minimum | Pull down slightly on the lift arms. |
position. | weight load to lower from the |
|
| uppermost position. |
|
NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare.
Stand Up Patient Lift | 20 | Part No. 1078984 |