Guidance and manufacturer’s declaration-electromagnetic immunity for all EQUIPMENT AND

SYSTEMS

Guidance and manufacturer’s declaration-electromagnetic immunity

The 2401LM Touchmonitor for Healthcare Applications is intended for use in the electromagnetic environment specified below. The customer, or the user of the 2401LM Touchmonitor for Healthcare Applications, should assure that it is used in such an environment.

Immunity Test

IEC 60601

Compliance Level

Level

 

 

 

 

 

Electrostatic

± 6 kV contact

± 6 kV contact

Discharge (ESD)

± 8 kV air

± 8 kV air

IEC 61000-4-2

 

 

 

 

 

Electrical Fast

± 2 kV for power

± 2 kV for power

Transient/Burst

supply lines

supply lines

IEC 61000-4-4

± 1 kV for

± 1 kV for input/output

 

input/output lines

Lines

 

 

 

Surge

± 1 kV line(s) to

± 1 kV line(s) to

IEC 61000-4-5

line(s)

line(s)

 

± 2 kV line(s) to

± 2 kV line(s) to earth

 

earth

 

 

 

 

Voltage Dips, Short

<5% UT

<5% UT

Interruption and

(>95% dip in UT)

(>95% dip in UT)

Voltage Variations

for 0.5 cycle

for 0.5 cycle

on Power Supply

40% UT

40% UT

Input Lines

 

(60% dip in UT)

(60 % dip in UT)

IEC 61000-4-11

for 5 cycles

for 5 cycles

70% UT

70% UT

 

 

(30% dip in UT)

(30% dip in UT)

 

for 25 cycles

for 25 cycles

 

<5% UT

<5% UT

 

(>95% dip in UT)

(>95% dip in UT)

 

for 250 cycles

for 250 cycles

 

 

 

Power Frequency

3 A/m

3 A/m

(50/60 Hz)

 

 

Magnetic Field

 

 

IEC 61000-4-8

 

 

 

 

 

Electromagnetic Environment - Guidelines

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.

Mains power quality should be that of a typical commercial or hospital environment.

Mains power quality should be that of a typical commercial or hospital environment.

Mains power quality should be that of a typical commercial or hospital environment. If the user of the 2401LM Touchmonitor for Healthcare Applications requires continued operation during power mains interruptions, It is recommended that the 2401LM Touchmonitor for Healthcare Applications be powered from an uninterruptible power supply or a battery.

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

NOTE UT is the A.C. mains voltage prior to application of the test level.

 

User Manual 2401LM

© 2013 Elo Touch Solutions, Inc. All rights reserved.

SW200127 Rev A - Page 8 of 32

Image 8
Contents Elo Touch Solutions 2401LM Touchmonitor SW200127 Rev aDisclaimer Trademark AcknowledgmentsSW200127 Rev a Page 3 Medical and Healthcare Application Disclaimer Classification Storag e / TransportationStorage / Transportation 10 % t o 90% Altitude O 3,000mEuropean Standards and Classifications Standards EN 60601-1-2Equipment and Systems IEC On Power SupplyElectrostatic Conducted RF Radiated RFIEC MHz to GHzTransmitter 100 Output PowerRated Maximum Table of Contents Product Description PrecautionsIntroduction Unpacking the Touch monitor InstallationConnector Panel & Interfaces Touchmonitor Connections Installing the Touch Technology Software Drivers SW200127 Rev a Page 16 Rear Vesa Mount MountingGeneral Mounting Information Ergotron Stand MountingVesa Mounting Options Power OperationTouch VideoOn-Screen Display OSD Temperature VolumePhase OSD TimeoutOSD and Power Lockouts AudioSolutions to Common Problems Problem Suggested TroubleshootingTechnical Support Technical AssistanceCare and Handling Safety & MaintenanceSafety Waste Electrical & Electronic Equipment Directive Weee Regulatory Information SW200127 Rev a Page 27 Ethers Pbde PcbaSW200127 Rev a Page 29 Warranty Information Agency Certifications Altitude Installing Software Drivers, 15 internal speakersIndex Getting in Touch with Us Check out our website