Guidance and manufacturer’s
SYSTEMS
Guidance and manufacturer’s
The 2401LM Touchmonitor for Healthcare Applications is intended for use in the electromagnetic environment specified below. The customer, or the user of the 2401LM Touchmonitor for Healthcare Applications, should assure that it is used in such an environment.
Immunity Test | IEC 60601 | Compliance Level | |
Level |
|
| |
|
|
| |
Electrostatic | ± 6 kV contact | ± 6 kV contact | |
Discharge (ESD) | ± 8 kV air | ± 8 kV air | |
IEC |
|
| |
|
|
| |
Electrical Fast | ± 2 kV for power | ± 2 kV for power | |
Transient/Burst | supply lines | supply lines | |
IEC | ± 1 kV for | ± 1 kV for input/output | |
| input/output lines | Lines | |
|
|
| |
Surge | ± 1 kV line(s) to | ± 1 kV line(s) to | |
IEC | line(s) | line(s) | |
| ± 2 kV line(s) to | ± 2 kV line(s) to earth | |
| earth |
| |
|
|
| |
Voltage Dips, Short | <5% UT | <5% UT | |
Interruption and | (>95% dip in UT) | (>95% dip in UT) | |
Voltage Variations | for 0.5 cycle | for 0.5 cycle | |
on Power Supply | 40% UT | 40% UT | |
Input Lines | |||
| (60% dip in UT) | (60 % dip in UT) | |
IEC | for 5 cycles | for 5 cycles | |
70% UT | 70% UT | ||
| |||
| (30% dip in UT) | (30% dip in UT) | |
| for 25 cycles | for 25 cycles | |
| <5% UT | <5% UT | |
| (>95% dip in UT) | (>95% dip in UT) | |
| for 250 cycles | for 250 cycles | |
|
|
| |
Power Frequency | 3 A/m | 3 A/m | |
(50/60 Hz) |
|
| |
Magnetic Field |
|
| |
IEC |
|
| |
|
|
|
Electromagnetic Environment - Guidelines
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the 2401LM Touchmonitor for Healthcare Applications requires continued operation during power mains interruptions, It is recommended that the 2401LM Touchmonitor for Healthcare Applications be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE UT is the A.C. mains voltage prior to application of the test level.
| User Manual 2401LM |