7CARE AND MAINTENANCE

Proper care and maintenance are essential to keeping your Lumex patient lift in a safe operating condition. In addition to inspecting the unit before each use, periodic maintenance checks should be done. We further recommend that this unit be serviced at least once annually by an authorized dealer.

WARNING: It is extremely important that the patient lift be inspected before each use. Ensure that all hardware and accessories are secure and that the actuator / motor is functioning properly. Failure to do so could result in patient / attendant personal injury or damage to your patient lift. Torn, cut, frayed or broken slings can fail, resulting in serious personal injury. Only use slings in good condition. Inspect before each use. Destroy and discard old, worn and unusable slings.

When you believe a component or part is not functioning properly, immediately contact your authorized Lumex dealer, as a potentially hazardous condition could exist.

BEFORE EACH USE

Before each use of the patient lift, ensure that:

All nuts and bolts are tight

The lift moves freely (with caster brakes off)

Rear caster brakes operate properly

Base adjustment mechanism operates easily

Spreader bar clips are secured tightly

The hydraulic pump operates properly (LF1031 only)

The electronic components operate properly (LF1041 only)

BATTERY

To increase the longevity of the battery, plug in the charger whenever possible (LF1041 only).

LUBRICATION

Your Lumex patient lift was designed with minimum maintenance in mind; however, a six (6) month check and lubrication should ensure that your patient lift will remain in peak operating condition. The casters MUST roll and swivel freely. Lightweight waterproof grease (such as automobile lubricant) should be applied to the ball bearings of the casters every six (6) months. Apply grease more frequently if the casters are exposed to periods of excess moisture. Lubricate all pivot points with lightweight oil. Clean patient lift of all excess oil and grease.

Before use of the patient lift, and every two months thereafter, place a drop of oil on the following areas:

The point where mast and boom connect

Spreader bar hooks

Pump handle hinge

Caster axles

The point where pump and boom connect (LF1031 only)

The point where actuator / motor and boom connect (LF1041 only)

LF1031-INS-LAB-RevD12

18

Image 18
Contents LF1031-INS-LAB-RevD12 GF Health Products, Inc Contents Introduction Safety Precautions LF1031-INS-LAB-RevD12 Features Unpacking Handling ProceduresShipping and Transportation Instructions InspectionAssembly Instructions ComponentsAssembly Base and mast assemblyLF1031-INS-LAB-RevD12 Power bracket Components LF1041 only attachingActuator / motor LF1041 only electrical power connections LF1041 only charging the battery LF1041 only control box power switch emergency stop buttonLF1041 only Emergency Procedures Operating InstructionsLift Operation Patient Transfer Lifting the patient from a bedLowering the patient onto a bed Lifting the patient from the floor Lowering the patient to the floorRemoving the sling Before Each USE Care and MaintenanceBattery LubricationCleaning Checking for Damage and WearCare of Slings LF1031 only Installation of 1010S005 Hydraulic Pump Removal of Existing PumpInstallation of New Pump Maintenance Schedule Troubleshooting LF1031 Hydraulic Patient Lift Symptom Possible Cause ActionSpecifications Limited Warranty GF Health Products, Inc. Graham-Field Index