8TROUBLESHOOTING
WARNING: If you experience a problem with your lift and are unable to service it yourself, contact your Lumex dealer.
LF1031 HYDRAULIC PATIENT LIFT
SYMPTOM | POSSIBLE CAUSE | ACTION |
Patient lift does not go up | Air in the cylinder | Bleed the cylinder by turning the release |
or down |
| knob and push down on the mast. Pump |
|
| up the lift again and turn the release knob |
|
| again. |
| Loss of hydraulic fluid (leaking | Replace any leaking cylinder |
| cylinder) |
|
Patient lift does not go | Air in the cylinder | See above Directions |
down |
|
|
Info: If the hydraulic cylinder begins leaking fluid, do not operate lift. Contact your Lumex dealer for a new cylinder.
LF1041 BATTERY POWERED PATIENT LIFT
SYMPTOM | POSSIBLE CAUSE | ACTION | |
Control box is not | Battery discharged | Charge battery | |
responding and relay is not |
|
| |
Battery charger defective | Replace battery charger | ||
clicking |
|
| |
Battery defective | Replace battery | ||
| |||
| Hand control pendant defective | Replace hand control pendant | |
Actuator / motor does not | Actuator plug not properly seated | Push actuator plug firmly into control box | |
run, but relay is clicking |
|
| |
Actuator defective | Replace actuator | ||
| |||
| Control box defective | Replace control box | |
Actuator / motor goes up | Hand control pendant defective | Replace hand control pendant | |
or down, but not both |
|
| |
Control box defective | Replace control box | ||
|