LCD Lift Codes

Listed below are codes that the lift controller outputs during lift operation. The codes will be displayed on an LCD screen located on the lift control board inside the pump module. See the Manual Operating Instructions in the operator's manual for pump cover removal instructions.

Non-Flashing Numbers

01 – Platform stowed

02 – Platform unfolding

03 – Platform unfolding paused

04 – Platform at floor level

05 – Platform beginning to lower

06 – Platform lowering (threshold cannot be occu- pied from this point down)

07 – Outer barrier moving to horizontal position

08 – Platform at ground level

09 – Outer barrier moving to vertical position

10 – Platform raising

11 – Platform raising paused at floor

12 – Platform folding (limited pressure)

13 – Platform folding (full pressure)

14 – Timed fold (cinching lift tite) or (anti-rattle state) 15 – Platform folding stopped

16 – Paused fold

17 – Platform between ground and 3” above ground

18 – Platform above 3”

19 – Outer barrier moving to horizontal postion

23 Outer barrier going back down after occupant detected

28– Illegal function/not defined

29– Interlock fault not recognized (or has been cleared but a motion button is still pressed)

30– Platform location unknown

31– Platform location transition state; attempting to locate position

35 – Two or more motion buttons are being pressed 36 – The retention belt cannot be buckled while try-

ing to fold or unfold

37 – Motion button being pressed is not a valid mo- tion

50 – Outer barrier is not up above inboard barrier locked position

51 – Threshold is occupied when platform is 1” or more below floor level

52 – Inner rollstop is not up and locked below inner rollstop locked position

53 – Inner rollstop occupied sensor is not activated between floor and inner rollstop up position

54 – Outer barrier is occupied before it is up

55 – Outer barrier is not latched when above the in- ner rollstop locked position (Millennium only)

56 – Outer barrier is not up and latched and bridge switch did not deactivate

57 – Outer barrier is not up and latched and ground detect switch did not deactivate (Century and Vista only)

58 – Outer barrier is not up and latched and the plat- form is 3” above the ground

59 – Outer barrier is not up after pausing platform travel

60 – Verify the IB occupied switch is functioning cor- rectly when lift is below floor level. The kickout gas springs are worn, replace before using

75 – Low voltage detected; must turn off power switch to reset LCD

77 – Vehicle secure interlock has not been activated 90 – Position will be set if you keep holding the button

until it beeps

91 – Position is out of a predetermined acceptable range of floor position

92 – Bridge switch is not made, needs adjusting

93 – Inner rollstop occupied switch is not made, position needs to be moved or switch should be adjusted

94 – Outer barrier is not made, fix and try again

95 – Outer barrier latch is not made (check for jumper on Century and Vista lifts, check latch on Millen- nium lifts)

99 – Controller program is not valid; replace controller

Flashing Numbers

Flashing 65 – Unfold button is pressed Flashing 66 – Fold button is pressed Flashing 67 – Down button is pressed Flashing 68 – Up button is pressed Flashing 69 – Bridge switch is activated Flashing 70 – Outer barrier latch switch is activated Flashing 71 – Ground detect switch is activated Flashing 72 – Outer barrier up switch is activated Flashing 73 – Inner rollstop up switch is activated Flashing 74 – Inner rollstop occupied switch is acti-

vated

Flashing 76 – Outboard barrier occupied switch is activated

Flashing 78 – Threshold tape switch “A” is activated Flashing 79 – Threshold tape switch “B” is activated Flashing 80 – Position set button is pressed

Page 11

Image 13
Contents CKA Congratulations Warranty and Registration InstructionsContents Lift Terminology Switch and Sensor Locations Certification Checklist Diagnostic Procedures Platform Fold Pressure Adjustment Millennium NL Series Platform Angle AdjustmentCentury NCL Series Vista NVL Series Platform Stop Blocks Platform Floor Level Adjustment Bridging Microswitch Bridging Microswitch AdjustmentAdjustment Guidelines LCD Lift Codes Troubleshooting ProceduresFlashing Numbers Non-Flashing NumbersLubrication Diagram Maintenance and LubricationCycles Maintenance and Lubrication Schedule1500 Cycles 1500 Consecutive 750Cycle Intervals Lift Electrical Schematic Unfold Lift Wiring DiagramDescription Symbol Hydraulic SchematicQty Description Hydraulics Parts ListHydraulics Diagram Part Numbers of Items Dedicated per Lift Model Pump Module Parts ListPump Mounting Bolts Pump Module DiagramRepair Parts List Exploded View This page intentionally left blank Providing Access to the World CKA