Leica MZ16 F user manual Additional components, Video/photo tube, Light source, Please read

Page 15

3.3 Additional components

￿Fasten the other components and other individ- ual accessories according to User Manual

M2-105-0:

Insert the binocular tube (2.5) or the video/ phototube into the dovetail ring on the FLUOIII® filter system and fasten it using a clamping screw.

Insert the eyepieces (4.6) in the tubes and fasten them using the clamping screws.

If necessary, place soft eyecups for prevent- ing eye infections over the built-in eyecups (see User Manual M2-105-0).

Screw the objective (4.7) into the bottom of the optics carrier (4.4) counterclockwise.

3.3.1 Video/photo tube

We recommend using the trinocular video/photo- tube (4.5) or the HD-V video/phototube with 100% light in the video/photo output, in order to achieve shorter exposure times at low fluorescence levels.

￿Insert the video/photo tube (4.5) into the ring connector on the FLUOIII® filter system accord- ing to User Manual M2-105-0 and fasten it using the clamping screw.

￿Fasten a binocular tube to the HD-V video/photo tube using a clamping screw.

3.4 Light source

The Leica 106 z lamp housing is the light source for the Leica MZ16 FA and is used with 100W or 50W mercury lamps, corresponding supply units and stray-light protection.

Please read

the separate User Manual M2-216-1 for the Leica lamp housing

the user manual and safety instructions for the ebq 100 supply unit

the user manual and safety instructions pro- vided by the lamp manufacturer, and, in partic- ular, the section on how to proceed if a lamp breaks and releases mercury.

Do not switch on the supply unit until you have read and familiarized yourself with the functions of the Leica MZ16 F, provided in the "Operation" chapter.

After it is switched on, the burner must be adjusted immediately.

3.4.1 Lamp housing 106 z

Fasten the 106 z lamp housing (4.8) to the adapter part as described in User Manual M2-216-1.

3.4.2 Supply units

The supply unit for 50W mercury lamps is described in User Manual M2-216-1.

The 100W mercury lamps are powered by the ebq 100 (4.9) supply unit.

Please read the separate instruction manual provided by the manufacturer.

Leica MZ16 F – Assembly

15

Image 15
Contents Leica MZ16 F Dear User Operating controls Contents Diagrams Safety conceptSafety instructions Symbols usedResponsibilities of the persons in charge of instrument Permitted usesNon-intended use Place of useLegal requirements Installation in third-party productsRepairs and servicing DisposalRisk of explosion, UV exposure, blinding Supply unit Light source safety instructionsPrecautions by the manufacturer Lamp housingOverview of the instrument Leica MZ16 F Overview of the instrument Assembly Assembly sequence Microscope components are best assembled in this order MZ16 FA Expansion diagramAssembly of stereomicroscope with stand Light source Lamp housing 106 z Supply unitsAdditional components Video/photo tubeLight spot see Equipping the filter changerCarrier Filter sets for fluorescenceInsert a total of four filter holders Simple filter holdersSlot for filter slide Inserting the filterFunctional principle Stereomicroscope settings and adjustmentsOperation Description Fluorescence technologyObservation without fluorescence Double iris diaphragmFilter sets Commissioning the fluorescence systemRapid filter changer Wait for 2 to 3 minutesWork flow Appendix Filter setsFilter sets Excitation filter Barrier filter Designation Technical data Optics carrier Leica MZ16 FFluorescence system Leica MZ16 F Leica MZ16 F Optical Data Optical accessories Leica MZ16 FStands, illuminators Leica MZ16 F Various accessoriesColumn with motorized focus Leica MZ16 F Technical Data WeightsFocusing drive, coarse/fine Maximum load 15 kg Ebq 100 supply unitObjectives Optical dataDimensions in mm DimensionsPage Leica Microsystems the brand for outstanding products