1.3 Light source: safety instructions

Precautions by the manufacturer

UV protection screen in front of the specimen plane prevents the user from looking directly into the UV rays.

Simple filter holders (that have only two open- ings for the observation beam paths) in the avail- able positions of the rapid filter changer prevent direct UV radiation from reaching the eyes.

UV shutter (which can be activated/deacti- vated) prevents direct UV radiation from reach- ing the eyes.

UV filters in the observation beam path protect the eyes.

The stray-light protection on the lamp housing prevents irradiation of the hands.

Warning

UV radiation can harm the eyes.

Therefore:

Never look into the light spot on the specimen plane without a UV protection screen.

Always fill all of the filter positions with filter holders.

Fill empty filter positions with the simple filter holders (which have only two openings for the observation beam paths).

Do not select a white, strongly reflective background for the stage.

Never open the lamp housing while the lamp is switched on.

Risk of explosion, UV exposure, blinding!

Supply unit

Always unplug the supply unit from the power supply:

when installing and disassembling the lamp housing

before opening the lamp housing

when replacing the high-pressure mercury burner and other parts, such as the heat- absorbing filter or the collector

during maintenance work on the supply unit.

Lamp housing

The Leica MZ16 F may be used only together with the Leica 106 z lamp housing for the 100W or 50W mercury burner, supply unit and stray-light protec- tion.

Leica Microsystems accepts no liability for damages resulting from the use of lamp housings from other manufacturers.

Never open the lamp housing while the lamp is switched on. Unplug the supply unit from the power supply:

Risk of explosion, UV exposure, blinding!

Before opening the lamp housing, allow it to cool off for at least 15 minutes.

Danger of explosion, danger of burns.

Never cover the air duct on the lamp housing.

Danger of fire!

Use the stray-light protection.

Danger to eyes from UV radiation.

Leica MZ16 F – Safety concept

9

Image 9
Contents Leica MZ16 F Dear User Operating controls Contents Diagrams Safety concept Safety instructions Symbols usedNon-intended use Permitted usesPlace of use Responsibilities of the persons in charge of instrumentRepairs and servicing Installation in third-party productsDisposal Legal requirementsPrecautions by the manufacturer Light source safety instructionsLamp housing Risk of explosion, UV exposure, blinding Supply unitOverview of the instrument Leica MZ16 F Overview of the instrument Assembly Assembly sequenceMicroscope components are best assembled in this order MZ16 FA Expansion diagramAssembly of stereomicroscope with stand Additional components Lamp housing 106 z Supply unitsVideo/photo tube Light sourceCarrier Equipping the filter changerFilter sets for fluorescence Light spot seeSlot for filter slide Simple filter holdersInserting the filter Insert a total of four filter holdersOperation Stereomicroscope settings and adjustmentsDescription Fluorescence technology Functional principleObservation without fluorescence Double iris diaphragmRapid filter changer Commissioning the fluorescence systemWait for 2 to 3 minutes Filter setsWork flow Appendix Filter setsFilter sets Excitation filter Barrier filter Designation Technical data Optics carrier Leica MZ16 FFluorescence system Leica MZ16 F Leica MZ16 F Optical Data Optical accessories Leica MZ16 FStands, illuminators Leica MZ16 F Various accessoriesFocusing drive, coarse/fine Maximum load 15 kg Leica MZ16 F Technical Data WeightsEbq 100 supply unit Column with motorized focusObjectives Optical dataDimensions in mm DimensionsPage Leica Microsystems the brand for outstanding products