5.2Commissioning the fluorescence system
After switching on the
Follow all safety instructions when handling mercury lamps and the lamp housing!
Switch on the supply unit (see User Manual
Wait for 2 to 3 minutes.
Adjust the burner in accordance with User Manual
For observation with fluorescence light, always position the aperture protection correctly to protect the observer (page 16).
2
1
Fig. 9 Filter changer, equipped with filter holders
1Filter sets
2Filter slide for filter holder with individual filters
5.3 Rapid filter changer
The rapid filter changer must be equipped with four filter holders at all times.
If you are using fewer than four filter hold- ers with fluorescence filters, always insert the included simple filter holders into the available positions. Otherwise there is a risk of danger to the eyes from direct UV radiation (see page 16).
If at all possible, avoid touching the filters in the filter sets in order to avoid fingerprints.
If you touch them, clean the filters using a soft,
Insert a filter set into the rapid filter changer such that the label (e.g. GFP) is upright and leg- ible on the right side.
•Ensure that the contour of the filter set matches the contour of the microscope.
Rotate the filter set by hand until it latches into place and insert the next filter set.
Insert a total of four filter sets or the included sim- ple filter holders into the empty positions.
20 | Leica MZ16 F – Functions |